background image

5

OPERATION

Illumination:

 On any light microscope, the ability to adjust light is probably

the most import element after the quality of the lens systems.  The Ken-A-Vision
Research Microscope has 4 distinct mechanisms to adjust light intensity, and
thereby allowing for High Bright Köhler Illumination.  The degree of light inten-
sity needed depends on individual conditions such as specimen
density, contrast, translucence  and staining, objective magnification and
individual users eyesight   Avoid turning the light onto full brightness as this
shortens the life of the bulb.

Too little or two much light is rarely recommended.  Light entering into the
objective from the sub-stage illuminator may be modified by:

Dimmer

 (Rheostat) switch on the bottom left side

A  Disc 

Iris

 diaphragm (

Field Diaphragm

) on the top of the illuminator 

in the base

A  Disc 

Iris

 diaphragm (

Condenser Diaphragm

) mounted within the 

Abbé  Condenser

Raising and lowering the condenser

The amount of light may also be manipulated by moving the gimbaled
condenser by using the three substage manipulating screws.

High Bright Köhler Illumination:

Köhler Illumination - Defined
Köhler illumination

 is a method to reduce an old problem of transmitted

light microscopes.  The filament of illuminating bulb of the microscope is often
visible in the same plane as the sample, reducing resolution and sharp focus.
One simple, partial solution is to reduce the amount of light striking the
specimen, however reducing voltage to reduce light intensity may also reduce
the breath of wavelengths (hues) reaching the specimen.  Various other
modifications have been done over time including opal bulbs, or placing a
glass diffuser lens in front of the bulb.  Though all these techniques work to
some degree, they tend to reduce the quality (some wavelengths of light are
lost) or the amount of incident light.  

Köhler illumination

 can be created

on microscopes with quality diaphragms, a moveable (both horizontally and
vertically) condenser, and a variable light source.  Uniformity of light is
essential to avoid shadows, glare, and inadequate contrast when taking
photomicrographs.

Köhler illumination

 overcomes these limitations.

Содержание Research Scope T-29031

Страница 1: ...cope T 29033 Binocular Acromat Plan Research Scope T 29043 Trinocular Acromat Plan Research Scope T 29034 Binocular Infinity Achromat Research Scope T 29044 Trinocular Infinity Achromat Research Scope T 29035 Binocular Infinity Semi Plan Research Scope T 29045 Trinocular Infinity Semi Plan Research Scope T 29036 Binocular Infinity Plan Research Scope T 29046 Trinocular Infinity Plan Research Scope...

Страница 2: ...2 ...

Страница 3: ...rs can operate the instrument conveniently and safely T 29031 T 29041 T 29032 T 29042 T 29033 T 29043 Halogen Light 12V 20 W D D D D D D High Bright Köhler Illumination D D D D D D Low Coaxial Fine Focus Scale 0 002mm D D D D D D 30 Inclined Binocular 55mm 75mm D D D 30 Inclined Trinocular 55mm 75mm D D D Eyepieces 10X D D D D D D DIN 4X 10X 40X 100X Oil Objectives Achromat Achomat Semi Plan Achom...

Страница 4: ...inocular 55mm 75mm D D D 30 Inclined Trinocular 55mm 75mm D D D Eyepieces 10X D D D D D D DIN 4X 10X 40X 100X Oil Objectives Achomat Infinity Achomat Infinity Semi Plan Infinity Semi Plan Infinity Plan Infinity Plan Infinity Double layered mechanical stage 180mm x 150 mm D D D D D D Gimbaled N A 1 25 Abbé Condenser with built in Iris Diaphragm and filters for Köhler Illumination D D D D D D ...

Страница 5: ...e manipulated by moving the gimbaled condenser by using the three substage manipulating screws High Bright Köhler Illumination Köhler Illumination Defined Köhler illumination is a method to reduce an old problem of transmitted light microscopes The filament of illuminating bulb of the microscope is often visible in the same plane as the sample reducing resolution and sharp focus One simple partial...

Страница 6: ...ntrast can introduce artifacts into your view and misrepresent what you actually see in the specimen Finally adjust the light intensity to your comfort level If available it is bet ter to use a neutral density filter in the filter holder below the bottom of the condenser then severely reducing the light intensity A neutral density filter tends to block all wave lengths equally Reduction of volt ag...

Страница 7: ...l a suitable position is found The stage should not move on its own as it sits waiting for your next use Bulb replacement Replacement bulb is Ken A Vision Part T6V20WP IMPORTANT NOTE Before replacing the lamp turn OFF the power switch and unplug from electrical source Place microscope on its side or back exposing the bottom of the unit Unscrew the knurled knob that is towards the front of the unit...

Страница 8: ... 100 of the field of view A Plan Achromat lens will also have the same corrections for chromatic aberrations as noted above In Plan Objectives when imaging the field parts of the specimen at the edge of the visual field are almost as well focused as those in the center Straight lines on the specimen appear straight throughout the visual field without any aberrations Such microscopes are especially...

Страница 9: ...lurry in places Spherical aberrations Spherical aberration occurs when the rays of light passing through the center of the lens tend to focus on a different plane than those entering near the edge of the lens This is because the curvature of the lens is not perfect The variance in cost of microscopes in general is often attributed to the cost of lenses needed to cure this problem Often the only wa...

Страница 10: ...avoid storing unit where it may be exposed to any chemical volatiles or other airborne corrosives Long Term Maintenance All optical and mechanical equipment requires periodic servicing to keep it performing properly Ken A Vision recommends the user anticipate this need by establishing a schedule of regular preventive maintenance to help to assure long life and sustain optimum performance of your i...

Страница 11: ...ssories If you are interested in purchasing accessories or replacement parts for your Research Scope visit the Ken A Vision Web site http store kenavision com catalog or contact your local Ken A Vision Dealer ...

Страница 12: ...en a vision com www ken a vision com INS SC29 Ken A Vision reserves the right to make design improvements and other changes in accordance with the latest technology There is no obligation to make changes in products already manufactured Patents Pending Copyright 2009 Ken A Vision Corporation ...

Отзывы: