background image

You have acquired one of the most unique and innovative changes in

microscopy in the last 50 years - A microscope which is capable of projecting

an XGA (resolution 1024 X 768) image onto a screen, interactive board or

even a wall.  Dependent upon the quality of your projector and the distance

the projector is from the screen or wall, you can magnify the image as much

as 300X and enlarge that image to greater than 5000 times original size!

G

Geettttiinngg  ssttaarrtteedd::

1. Plug X2000 into your power source, using the Ken-A-Vision power 

supply provided with the unit.

2. Connect a VGA cable (not provided by Ken-A-Vision) into the VGA 

plug at the rear of the X2000, and then plug the VGA connector into 

the VGA input of your projector. 

3. Turn on X2000 with the switch at the rear of the unit and turn on 

your projector, being sure correct button is activated.

4.  If  possible,  be  sure  that  your  projector  is  sharply  focused,  by 

utilizing any text or symbols that can be activated using the menu 

button on the projector.

5. What ever the camera of the X2000 sees at that point should be 

visible on the screen, and if you have centered correctly it should be 

some  part  of  the  slide  or  Petri  dish  (hereafter  referred  to  as 

"specimen") you have placed on the stage.

PPllaacciinngg  tthhiinnggss  oonn  tthhee  ssttaaggee::

The X2000 is provided with a floating stage and mechanical stage. A

floating stage allows you to reposition the stage by grasping the side of the

stage using your thumb and fingers, and moving the stage to any position you

wish. Center the slide or object you wish to view with the X2000 by easy to use

mechanical stage installed on the unit.

Place the slide (specimen) in the mechanical stage , by pulling back on

the retractable spring arm, placing the slide against the straight edge at the

right side (as you face the front of the X2000 Microprojector), and then gen-

tly releasing the spring loaded retractable arm so that the specimen is locked

into place.

Now the specimen may be manipulated into place, by either moving the

floating stage, or using the knurled controls of the Mechanical Stage or both.

LLiigghhtt  SSoouurrccee::

Light for our X2000 is provided by a variable light source at the top of

the microscope.  Turning the orange dial will increase or decrease the amount

of light hitting the stage, as indicated by bar meter.

TThhee  ZZoooom

m  O

Obbjjeeccttiivvee  LLeennss::

The Zoom Objective Lens is located below the stage.  On the front side

(as you face the front of the X2000 Microprojector) of the tube of the Zoom

Object Lens, there is an arrow pointer to set each point.  On the moveable

part of the Zoom Objective Lens are a series of numbers (3X, 4.5X, 7.5X, 11X,

16X,  20X,  26X,  30X)    indicating  the  approximate  magnification  achieved

when the Zoom Objective Lens is in that position.  Turning the moveable por-

tion of the Zoom Objective Lens counter-clockwise increases the magnifica-

tion from approximately 3X to 30X.  In doing so the magnification achieved is

approximately going from 30X to 300X.  

BBeesstt  PPrraaccttiiccee  FFooccuussiinngg  --  ZZoooom

miinngg::

1. Place the specimen to be observed on the stage, and center as well 

as possible with your naked eye.

2. Be sure that the Zoom Objective Lens is set on 3X, with the light 

dimmer set to minimum and included filter is screwed in)  Note: the 

filter  is  designed  to  be  used  between  3X-4.5X  only.  It  MUST  BE 

REMOVED above 4.5X magnification.

3. Using the Coarse and Fine Adjust knobs and observing the picture 

on the screen, bring the specimen into focus, similarly to the way one 

focuses a standard microscope.  (Be sure that your projector's focus 

previously was sharply set for the conditions under which you are 

using it).

4. Change the light impinging on the specimen using the orange dial 

at the top of the X2000. Please note that often objects are better 

viewed in less light then at maximum incident light, so experiment 

with the light at 4X to see if there is an optimum light point.

5.  Now  slowly  turn  the  Zoom  Objective  Lens  counter-clockwise, 

increasing the magnification as you do this.  You may have to read

just the focus of the X2000 Microprojector using the fine and coarse 

adjust knobs. 

6. As you increase magnification, you may also have to increase the 

intensity  of  the  light,  as  just  like  regular  microscopes,  as  the 

magnification poser of an objective lens increase, the amount of light 

necessary must also increase.

Отзывы: