6. Maintenance
32
6-2. Periodic Check
Periodic check is necessary to verify performance of refractometer. When unexpected
physical force is given to the unit, precision of measurement must also be checked to see
if different results should appear.
Check by user
1)
Check or calibrate the unit according to the instruction in manual.
2)
Measure Refractive index standard liquid or Brix converted RI standard liquid. For
check reference solution, it is recommended to use a standard liquid of which refractive
index is close to the value that you frequently obtain in measurement.
(RI standard liquid and Brix converted RI standard liquid are available from KEM.
See [6-4. Consumables and parts])
3)
If repeated measurements result with an error deviated from range in [8-3.
Specifications], consult with your local dealer or sales representative.
Customer service contract (paid service)
Your refractometer is checked and serviced according to ISO 9001 Quality Management
System. Either your unit is serviced at your site or sent back to KEM service.
Our service work is conducted using measuring equipment which is traceable to national
or international standard of metrology. Certificate of Calibration is issued after service
work is complete and the unit is returned to user.
For technical assistance and request for service work, contact your local dealer or sales
representative.
Содержание RA-600