Keller LEO1-Ei Скачать руководство пользователя страница 1

Eigensicheres digitales Manometer mit Spitzendruckwert-Erfassung und Min.-/Max.-Anzeige.

Intrinsically Safe Manometer with Peak Pressure Value Detection and Min.-/Max.-Display. 

Manomètre numérique de sécurité intrinsèque avec mesure de pics et affichage Min./Max..

LEO1-Ei

zum Einsatz in explosionsgefährdeten Bereichen

for use in Hazardous Applications

pour utilisation en zones explosibles

Min.-/Max.-Druckwert,

Spitzenwert
Min.-/Max. Pressure Value,

Peak Value
Valeur de pression Min./Max.,

Pic de pression

Aktueller Druckwert

Actual Pressure Value

Valeur de pression actuelle

SELECT

ENTER 

Min./Max.

Druckanschluss

Pressure Connection

Raccord pression

Содержание LEO1-Ei

Страница 1: ...c mesure de pics et affichage Min Max LEO1 Ei zum Einsatz in explosionsgef hrdeten Bereichen for use in Hazardous Applications pour utilisation en zones explosibles Min Max Druckwert Spitzenwert Min M...

Страница 2: ...es the selected function or pressure unit The right key is also used to switch between the Min and Max pressure value Turn on Pressing the SELECT key turns the instrument on The instru ment first disp...

Страница 3: ...cm2 Example Setting a new Zero Reference Turn on the instrument by shortly pressing SELECT Wait for the instrument s measuring mode 3 s Press the SELECT key 3 times MANO appears Press ENTER PEAK on or...

Страница 4: ...or CONT on function is activated it is in dicated with a flashing sign on the display OFF flashes when CONT on is set 5 If a pressure can not be represented on the display OFL overflow or UFL underfl...

Страница 5: ...ery starts weak ening a low battery warning BAT LOW will appear in the display Battery change Please turn off the instrument before changing the battery Open the instrument by turning the display ring...

Страница 6: ...Calibration The ZERO function allows to set any pressure value as a zero reference The factory setting of the pres sure zero for the ranges 61 bar absolute is at vacuum 0 bar absolute For relative pr...

Страница 7: ...ra tus it can be operated in explo sive atmospheres The operating ambient tempe rature are inclu ded between 10 C and 80 C Connect the pressure port of the digital Manometer to neutral earth to ground...

Страница 8: ...the following norms for this Intrinsically Safe Mano meter LEO1 Ei are applied and the EC examination certifi cate under LCIE 01 ATEX 6001 X Issue 04 is given This declaration is given for the manufac...