Keller HP Series Скачать руководство пользователя страница 8

5

m

ano

-m

ode

The M

ano

-Mode is the standard mode of the ca-

librator.  In  this  mode,  the  calibrator  shows  on 
the upper display the actual pressure generated. 
The  smaller  displays  below  indicate  the  peak- 
and trough values. The measured values on the 
smaller displays always appear with a reduced 
resolution. Activating e

xe

 resets the peak- and 

trough values, meaning that they are reset to the 
actual pressure.

Full-Version only: If a transmitter is connected, 
the  display  automatically  indicates  the  value 
of  the  transmitter  in  mA  or  V  on  the  lower  left 
display. The lower right display shows the tem-
perature.

Pressing  the  H

old

-Key  (H-Key)  freezes  the 

measured  values.  The  actual  pressure  is  now 
shown on the lower right display. Pressing the 
H-Key  again  leads  back  into  the  M

ano

-Mode 

with  the  peak-  and  trough  values  reset  to  the 
actual pressure (same as e

xe

).

m

ano

-m

ode

 

d

isPlay

 

WitHout

 t

ransMitter

 

= Is being read as a “1” between 100 bar and 200 bar when displaying with increased resolution --> 133.41 bar

Actual 

Pressure

Trough

Peak

A

cti

vate E

xE

 

Actual 

Pressure

Trough

Peak

Press H-Key

H

old

-m

ode

 

Frozen 

Value

Actual Pressure

Peak

With H-Key back to 
M

Ano

-Mode

m

ano

-m

ode

 

d

isPlay

 

WitH

 t

ransMitter

 

(Full-Version only)

Actual 

Pressure

Temperature of 

the Reference-

Sensor

Signal of

Transmitter

Press H-Key

H

old

-m

ode

 

Frozen

Value

Actual 

Pressure

Frozen Value

of Transmitter

Содержание HP Series

Страница 1: ...High Pressure Calibrator Standard and Full Version KELLER May 2006...

Страница 2: ...ble RS 232 2 Program Discs Logger Windows Logger MS DOS Included in Delivery High Pressure Calibrator Full Version 1 Calibrator 1 Carrying Case 1 Test Unit Adapter G 1 4 G 1 2 1 Test Unit Adapter G 1...

Страница 3: ...ing of Pressure Switches 9 Pst Mode Preparing Storage of Measured Values 11 Rec Mode Executing Storage of Measured Values 12 Rec Mode Measuring the Linearity 13 Lin Mode Calculating the Linearity 14 L...

Страница 4: ...Refill Screw 10 Oil Screw Compressor 11a Outlet Pressure Port G 1 4 female with Hexagon Plug 200 bar and 350 bar Calibrator with Overpressure Valve 11b Outlet Pressure Port G 1 4 female with Hexagon P...

Страница 5: ...essure with both the integrated priming pump and the oil screw compressor The pressure can be fine adjusted The high technology integrated in this instru ment allows it to measure precisely both the p...

Страница 6: ...within the oil chamber the oil refill screw 9 must be open when operating the calibrator Generate pressure with the priming pump 12 up to approx 10 bar Increase or reduce pressure with screw com press...

Страница 7: ...Mano Release the F Key Within specific functions the F Key can execute additional commands Exe Step Reset Activating Off turns off the instrument When starting the instrument again the display will au...

Страница 8: ...ay shows the tem perature Pressing the Hold Key H Key freezes the measured values The actual pressure is now shown on the lower right display Pressing the H Key again leads back into the Mano Mode wit...

Страница 9: ...e With the command Exe the new zero is being determined and stays as a reference until a new Tara is being executed or until the instrument is turned off The program returns to the Mano Mode The Unit...

Страница 10: ...sure Activate Sel Flashing Exe changes to hr hours Exe again to day Activate Step First digit is flashing Exe encreases digit With Step to next digit Activate Step Second digit is flashing Exe encreas...

Страница 11: ...o return to the Prep Mode from the End display A new test can be started as follows also when measurements are running Press the H Key or Activate Run with the F Key At any time you may return to the...

Страница 12: ...ing recorded and can be written into a PC and displayed on the screen only with Logger Option Activating End followed by the activation of another function quits the Pst Mode and leads into the new se...

Страница 13: ...Low if the measurement started at high pressure and the pressure switch was switching at a lower pressure An increase of pressure from this status writes the new switch point automatically into the sm...

Страница 14: ...er recording the storage requirements are tripling 1000 measurements with three values each lead to 3000 measuring values Adjustment of the Storage Parameters Activate Rec Recording interval Number of...

Страница 15: ...ay At any time a record can prematurely be termi nated by activating Res In the Rec Mode you may also display the tem perature Pressing the H key at Prep display indicates the temperature on ther lowe...

Страница 16: ...Maximal 10 linearity values are allowed After having activated Lin activating the com mand Save writes the actual values of the reference pressure and of the test unit into the memory This cycle can...

Страница 17: ...ersion only From this position the linearity of all other measuring points can be called upon Pressure at point 3 Activate Exe Pressure at point 4 14 Number of last measuring point Activate Calc Actua...

Страница 18: ...endency of posi tion or the influence of temperature can result in zero shifts These shifts can be permanently corrected as follows Activate Zero Actual Zero Activate Exe Activate Reso Activate Reso T...

Страница 19: ...e absolute time axis is calculated from the time set in the PC the time meter of the calibrator and the recorded time The calibrator should therefore not be turned off between the recor ding and the t...

Страница 20: ...necting your test unit Open the battery compartment of the instrument base and change the batteries Restarting the Calibrator If it should occur that the programme is locked meaning that the instrumen...

Страница 21: ...transmitter in mA or V Connecting Pressure Switches Plug in the pressure switch adapter Internal supply is OK when LED 3 lights Connect the pressure switch to adapter FUNCTIONS Mano Display of pressu...

Страница 22: ...otal Oil Chamber Volume Oil Type Dead Volume Total Weight including Batteries Dimensions L x W x H Protection only Standard 0 1 FS Full Scale from 0 50 C 0 1 FS 1 Digit 10 mbar 100 mbar 100 mbar FS 10...

Страница 23: ...ion of which a drawing marked with the CE sign exists and which are produced according to this drawing This explanation has no validity without such a drawing As criteria for the electromagnetic compa...

Страница 24: ...www keller druck com...

Отзывы: