The calibrator is turned on by pressing the func-
tion key once.

c

onnectIng

 t

est

-u

nIts

*

Adapter:
Included  in  the  delivery  of  the  calibrator  are 
 different test-unit adapters, one for each of the 
following transmitter types:
–  4…20 mA / 0…20 mA transmitter
–  0…10 V / 0…20 V transmitter
–  Pressure switch

Plug the test-unit adapter into socket 14.

e

lectrIcal

 c

onnectIon

 

oF

 t

est

-u

nIt

*

Select the correct adapter. Wire the transmitter/ 
switch  under  test  to  the  adapter.  Check  that 
 polarity is observed and that good connections 
are made.

P

ressure

 c

onnectIon

 

For

 

tHe

 t

est

-u

nIt

**

Your  test-unit  can  be  connected  with  the 
 calibrator via the pressure connection 11a resp. 
11b.

P

ressure

 g

eneratIon

Attention:  To  avoid  an  overpressure  within  the 

oil chamber, the oil refill screw (9) must be open 

when operating the calibrator!

Generate  pressure  with  the  priming  pump  (12) 
up to approx. 10 bar.
Increase  or  reduce  pressure  with  screw  com-
pressor (10).

d

ecreasIng

 

tHe

 P

ressure

1)  To return pressure to zero, 

completely 

  

 

screw back compressor (10).

2)  Open shut-off valve (5).

Do  not  open  (5)  when  system  is  under  high 
pressure! 

(Especially  when  using  an  external 

pres sure source, damage/injury may result).

If you are not able to reach the desired pressure, 
read  the  chapter  “Maintenance”  on  page  17  on 
how to completely prime the pressure system.

e

xecutIng

 F

unctIons

 

and

 c

ommands

Read  the  following  chapter  to  get  acquainted 
with the functions.

t

urn

-o

FF

Press  the  function  key  until  the  command  o

FF

 

appears on the display. Release the key imme-
diately and the instrument turns off.

**  To  avoid  leakage  and  damage  of  the  check  valve,  the  test-unit  has  to  be  cleaned  prior  to  connecting  it  to  the 

 calibrator.

Содержание HP Series

Страница 1: ...High Pressure Calibrator Standard and Full Version KELLER May 2006...

Страница 2: ...ble RS 232 2 Program Discs Logger Windows Logger MS DOS Included in Delivery High Pressure Calibrator Full Version 1 Calibrator 1 Carrying Case 1 Test Unit Adapter G 1 4 G 1 2 1 Test Unit Adapter G 1...

Страница 3: ...ing of Pressure Switches 9 Pst Mode Preparing Storage of Measured Values 11 Rec Mode Executing Storage of Measured Values 12 Rec Mode Measuring the Linearity 13 Lin Mode Calculating the Linearity 14 L...

Страница 4: ...Refill Screw 10 Oil Screw Compressor 11a Outlet Pressure Port G 1 4 female with Hexagon Plug 200 bar and 350 bar Calibrator with Overpressure Valve 11b Outlet Pressure Port G 1 4 female with Hexagon P...

Страница 5: ...essure with both the integrated priming pump and the oil screw compressor The pressure can be fine adjusted The high technology integrated in this instru ment allows it to measure precisely both the p...

Страница 6: ...within the oil chamber the oil refill screw 9 must be open when operating the calibrator Generate pressure with the priming pump 12 up to approx 10 bar Increase or reduce pressure with screw com press...

Страница 7: ...Mano Release the F Key Within specific functions the F Key can execute additional commands Exe Step Reset Activating Off turns off the instrument When starting the instrument again the display will au...

Страница 8: ...ay shows the tem perature Pressing the Hold Key H Key freezes the measured values The actual pressure is now shown on the lower right display Pressing the H Key again leads back into the Mano Mode wit...

Страница 9: ...e With the command Exe the new zero is being determined and stays as a reference until a new Tara is being executed or until the instrument is turned off The program returns to the Mano Mode The Unit...

Страница 10: ...sure Activate Sel Flashing Exe changes to hr hours Exe again to day Activate Step First digit is flashing Exe encreases digit With Step to next digit Activate Step Second digit is flashing Exe encreas...

Страница 11: ...o return to the Prep Mode from the End display A new test can be started as follows also when measurements are running Press the H Key or Activate Run with the F Key At any time you may return to the...

Страница 12: ...ing recorded and can be written into a PC and displayed on the screen only with Logger Option Activating End followed by the activation of another function quits the Pst Mode and leads into the new se...

Страница 13: ...Low if the measurement started at high pressure and the pressure switch was switching at a lower pressure An increase of pressure from this status writes the new switch point automatically into the sm...

Страница 14: ...er recording the storage requirements are tripling 1000 measurements with three values each lead to 3000 measuring values Adjustment of the Storage Parameters Activate Rec Recording interval Number of...

Страница 15: ...ay At any time a record can prematurely be termi nated by activating Res In the Rec Mode you may also display the tem perature Pressing the H key at Prep display indicates the temperature on ther lowe...

Страница 16: ...Maximal 10 linearity values are allowed After having activated Lin activating the com mand Save writes the actual values of the reference pressure and of the test unit into the memory This cycle can...

Страница 17: ...ersion only From this position the linearity of all other measuring points can be called upon Pressure at point 3 Activate Exe Pressure at point 4 14 Number of last measuring point Activate Calc Actua...

Страница 18: ...endency of posi tion or the influence of temperature can result in zero shifts These shifts can be permanently corrected as follows Activate Zero Actual Zero Activate Exe Activate Reso Activate Reso T...

Страница 19: ...e absolute time axis is calculated from the time set in the PC the time meter of the calibrator and the recorded time The calibrator should therefore not be turned off between the recor ding and the t...

Страница 20: ...necting your test unit Open the battery compartment of the instrument base and change the batteries Restarting the Calibrator If it should occur that the programme is locked meaning that the instrumen...

Страница 21: ...transmitter in mA or V Connecting Pressure Switches Plug in the pressure switch adapter Internal supply is OK when LED 3 lights Connect the pressure switch to adapter FUNCTIONS Mano Display of pressu...

Страница 22: ...otal Oil Chamber Volume Oil Type Dead Volume Total Weight including Batteries Dimensions L x W x H Protection only Standard 0 1 FS Full Scale from 0 50 C 0 1 FS 1 Digit 10 mbar 100 mbar 100 mbar FS 10...

Страница 23: ...ion of which a drawing marked with the CE sign exists and which are produced according to this drawing This explanation has no validity without such a drawing As criteria for the electromagnetic compa...

Страница 24: ...www keller druck com...

Отзывы: