Short Form Manual EV-94 EB
This short form manual allows a quick basic configuration. For more details
please refer to the "Manual for connection and operation of the EV-94 B".
Note: If no key is pressed for 30 seconds, the configuration must be reset!
Action
Display
Comment
Example
Start-up by connection to current voltage
xxx....
Depending on configuration
Err.x, 0, xxx
Press button 2 for > 2 Sekunden
Hi, ..x.. Inp
Input signal
Press
button 2 or 3
U, I, t.rES usw.
I for current signal (0..20mA, 4..20mA)
I (for current measurement)
“
button 1
InP
Acknowledgement of input signal
“
button 1
SEnS
Signal
“
button 2 or 3
4..20, 0..20
Select signal
4..20 ( for 4..20mA)
“
button 1
SEnS
Acknowledge signal
Press
button 1
dp
Select position of decimal point
“
button 2 or 3
- - - -
Select position of decimal point
i.e.
-. - - -
“
button 1
dp
Acknowledge position of decimal point
“
button 1
di.Lo
Low input value (Display Low)
“
button 2 or 3
0000
Select low input value
0.000
“
button 1
di.Lo
Acknowledge low input value
“
button 1
di.Hi
High input value (Display High)
“
button 2 or 3
2.500
Select high input value
2.500
“
button 1
di.Hi
Acknowledge high input value
Press
button 1
Li
Measuring range limit (Limit)
“
button 2 or 3
off, on.Er, on.rG
Select measuring range limit
off
“
button 1
Li
Acknowledge measuring range limit
“
button 1
FiLt
Filter (Filter value 0.01 ...2.00 seconds)
“
button 2 or 3
0.00
Select filter
1.00
“
button 1
FiLt
Acknowledge filter
Press
button 1
outP
Output (switching point, configuration)
“
button 2 or 3
2P
Select output, 2-level
2P
“
button 1
outP
Acknowledge output
“
button 1
1.dEL
Delay (switching output 1, delay)
“
button 2 or 3
1.00
Select delay
1.00
“
button 1
1.dEL
Acknowledge delay
Press
button 1
1.out
Type switching output (NPN, PNP, R+)
“
button 2 or 3
nPn, PnP, Pu.Pu
Select switching output type
nPn
“
button 1
1.out
Acknowledge switching output type
“
button 1
1.Err
Switching output if error (active, inactive)
“
button 2 or 3
off, on
Select switching output if error
off
“
button 1
1.Err
Acknowledge switching output if error
Press
button 1
1.on
Switching output 1, On
“
button 2 or 3
0000
Switching output 1 On > = 0.505
0.505
“
button 1
1.on
Acknowledge switching output 1
“
button 1
1.oFF
Switching output 1, Off
“
button 2 or 3
0000
Switching output 1 < = 0.495
0.495
“
button 1
1.oFF
Acknowledge switching output 1
“
button 1
1.oFF
End