Operating instructions CellaTemp PKx 6x
34
5
Confirm parameter value
►
Press [Enter]
> The display indicates the
parameter� The new value has
been saved and will take effect�
Adjust additional parameters
►
Start again with step 2�
Exit operating parameters layer
►
Wait 30 seconds
or
► Press [▲] or [▼] to change to the paramter
End.
Then press [Enter] to
change to the functional menu
► In the functional menu press [▲] or [▼] to change to the parameter
End
,
then press [Enter]�
The instrument features a keylock� Activate/deactivate the keylock as
follows:
► Press key [▲▼] simultaneously and hold them down for 10 sec.
> The display shows Loc or uLoc for 1 second to signalize the change�
If you press both keys [▲▼] only briefly, you will exit the layer
(ESC function)�
13.2 Damping function
When the target object‘s temperature is erratic, the damping function smoothens
these temperature fluctuations in order to stabilize the measuring signal� The
greater the time constant [
DAP
], the lower the effect of these fluctuations on the
yielded temperature reading�