background image

Pyrometer 

CellaTemp PKL 63, 

PKF 66/67, PK(L) 68

Ident no.: 1086021   06/2017

Содержание CellaTemp PKL63

Страница 1: ...Pyrometer CellaTemp PKL63 PKF66 67 PK L 68 Ident no 1086021 06 2017...

Страница 2: ...8 15 Function 9 5 1 Analogue output 9 5 2 Switching threshold OUT1 10 5 3 Output signal 10 5 4 Upper threshold delay 10 5 5 Lower threshold delay 11 5 6 Switching functions 11 5 7 Internal signal pro...

Страница 3: ...g threshold OUT1 25 10 3 Switching threshold OUT2 26 10 4 Measurement acquisition two colour mode 27 10 5 Measurement acquisition one colour mode 28 10 6 Advanced Features 28 11 Setup 29 11 1 Two colo...

Страница 4: ...5 4 Measurements errors 41 16 Ways to determine emissivity 41 16 1 Emissivity coefficient tables One colour mode 42 17 Maintenance 43 17 1 Cleaning the pyrometer lens 43 18 Shipping packaging and disp...

Страница 5: ...ATTENTION This symbol points out guidelines If you do not observe them the device might be damaged malfunctioning or even fail to operate CAUTION This symbol points out hints and information which sho...

Страница 6: ...n this manual Operational safety can only be ensured when the instrument is used for its intended purpose The use of the pyrometer for any other purpose beyond what is specified in this manual is proh...

Страница 7: ...y a hot object and converts this to an electric switch signal The advantage of this technique is that there is no mechanical contact between the sensor and the hot object The instrument is suitable fo...

Страница 8: ...s screens or viewing windows that reduce energy and by dirt smoke or steam in the atmosphere The two color mode can also be used on targets that do not com pletely fill the target spot provided the ba...

Страница 9: ...utput can be used The following sources are available Two colour mode Ao Ao S QUot One colour mode Ao Ao S SpEc Within the overall measuring range the required measuring range can be set to C or F usi...

Страница 10: ...d as d1 rP would then fall below its minimum value the d1 rP is kept with its minimum value If d1 SP subsequently increases again d1 rP also immediately increas es The minimum distance between d1 SP a...

Страница 11: ...g by a downstream control system Lower threshold delay d1 d1 dr 0 10 sec 5 6 Switching functions 1 Temperature profile 2 Switch signal Hno 3 Switch signal Hno with upper and lower threshold delay 4 Sw...

Страница 12: ...ctive glass or the sighting cone of the pyrometer gets dirty during a measurement The dirt alert is activated with parameter d2 d2 s DirT The dirt alert is a warning function This warning will become...

Страница 13: ...the internal temperature is 70 C the switching output OUT 2 switches back to its original state 5 12 Output signal The following switching functions can be selected for dirt alert signal strength and...

Страница 14: ...vity coefficient continues to be adjusted in the background This is an easy way to determine the emissivity when the object temperature is known The changed values are directly adopted 5 16 Emissivity...

Страница 15: ...he transmis sion properties of any supplementary lens and or protective window screwed onto the pyrometer Set the pyrometer for the specific transmission factor a percentage value which is indicated e...

Страница 16: ...ent according to the following schematic 2 3 1 4 5 1 4 L L 24VDC 5 C Q 2 0 4 20mA Out2 3 shielded Pin 1 BN brown L Power supply 24V DC Pin 4 BK black Open Collector switching output Imax 150 mAOUT1 or...

Страница 17: ...lising current to flow between devices through a cable shielded at both ends In this case be sure to install an additional equipotential bonding line To avoid an equalising current the pyrometer can b...

Страница 18: ...easured temperature 1 to 3 Indicator LEDs LED 1 indicates switching output OUT1 LED 2 indicates switching output OUT2 LED 3 signal strength 4 Control key Enter Select parameter and confirm setting 5 C...

Страница 19: ...19 Operating instructions CellaTemp PKx 6x EN 9 Menu 9 1 Process value display lock unlock display shows Loc uLoc 9 2 Analogue output Menu is not available when key lock is active Display Adjustment...

Страница 20: ...Operating instructions CellaTemp PKx 6x 20 9 3 Switching threshold OUT1 Menu is not available when key lock is active Display Adjustment...

Страница 21: ...21 Operating instructions CellaTemp PKx 6x EN 9 4 Switching threshold OUT2...

Страница 22: ...Operating instructions CellaTemp PKx 6x 22 9 5 Two colour mode Menu is not available when key lock is active Display Adjustment...

Страница 23: ...23 Operating instructions CellaTemp PKx 6x EN 9 6 One colour mode Menu is not available when key lock is active Display Adjustment...

Страница 24: ...Operating instructions CellaTemp PKx 6x 24 9 7 Advanced Features Menu is not available when key lock is active Display Adjustment...

Страница 25: ...lue for scaling Ao 04 Analogue output 0 4 20 mA 0 20 mA scaling Analogue output 4 20 mA scaling Analogue output End End 10 2 Switching threshold OUT1 Parameter Function Explanation d1 S OUT1 source QU...

Страница 26: ...re Dirt dirt alert Sync sync pulse d2 SP OUT2 Upper threshold Upper threshold which activates OUT2 d2 rP OUT2 Lower threshold Lower threshold which activates OUT2 D1 Fn Output function no normallly cl...

Страница 27: ...13 4 Q Ano Display mode during sampling time T 0 show lower limit of temp range during running measurement t hld Hold previous temp reading during running measurement Q dAP Damping after min max memo...

Страница 28: ...0 1 sec steps S dAP Damping after min max memory Off off Time in sec in 0 1 sec steps S tMP One colour temperature reading One colour temperature reading End End Menu item only if Peakhold is activate...

Страница 29: ...is calculated for this signal LED 3 lights up Measurement OK LED 3 flashes Signal strength below the set threshold measurement is critical LED 3 off The signal strength is too low a measurement is not...

Страница 30: ...5 Emissivity coefficient tables An incorrectly set emissivity coefficient leads to wrong temperature readings Set the emissivity coefficient after connecting the supply voltage or resetting the parame...

Страница 31: ...For focal adjustment loosen the shown socket screw hexagon socket screw DIN 916 with a wrench DIN 911 and shift the internal body of the tube towards the lens tube Due to the O ring sealing between th...

Страница 32: ...Do not leave the instrument unattended when the laser is activated Do not point the laser beam at any person During pyrometer installation and alignment make sure to avoid the possibility of laser lig...

Страница 33: ...meter Press Enter to access the main menu 2 Select output function Press key until the required output function or the Advanced Features is displayed 3 Show parameter value Press Enter Current paramet...

Страница 34: ...ge to the parameter End then press Enter The instrument features a keylock Activate deactivate the keylock as follows Press key simultaneously and hold them down for 10 sec The display shows Loc or uL...

Страница 35: ...ing in temperature readings which would appear to be cyclical In this mode the displayed temperature reading will not drop between targeted objects The peak temperature reading will be held for a pres...

Страница 36: ...al flow For example thus can determine the temperature of bolts with variable lengths that move acyclic along the pyrometer A temperature threshold has to be defined for this purpose The measurement s...

Страница 37: ...ng the measurement the previous read ing is available at the temperature output d2 q Ano t 0 During the measurement the temperature is set to the lower limit of the temperature range 1 Measuring objec...

Страница 38: ...10 Setup to obtain exact measuring results 14 Operation After connecting the supply voltage the pyrometer will be automatically initialized and will perform a self diagnosis After approx 0 5 sec the s...

Страница 39: ...surements All materials radiate thermal energy in all states of aggregation above absolute zero This radiation is mainly caused by atomic or molecular oscillations This temperature radiation is only a...

Страница 40: ...body emits at any wavelength the maximum energy possible for the specific temperature Real bodies do not have this ability In other words a black body completely absorbs the radiation without reflecti...

Страница 41: ...ed These objects then have to be shaded This effect is particularly to be observed in the measurement of a colder object within a hot oven 16 Ways to determine emissivity Technical literature or opera...

Страница 42: ...s the previous measurement reading again 16 1 Emissivity Table One colour mode Overview of the emissivities of different materials in Model PKx 6x Wavelength 0 8 1 1 m Black Body 100 Aluminium polishe...

Страница 43: ...package shows visible signs of damage please refuse the shipment If this is not possible accept the shipment on the condition that the freight carrier s delivery record is noted with the extent of the...

Страница 44: ...2 Cooling jacket PK 01 B AF 1 1067753 Bayonet coupling PS 11 N AF 4 561585 90 Deflection mirror PS 11 W 561955 Supplementary tube ZA 01 Q 35 514234 Intermediate tube ZA 01 M 513807 Intermediate tube Z...

Страница 45: ...ission rate COM2 38 400 Baud Storage temperature 20 80 C Permissible humidity 95 r H max non condensing Power requirement 24 V DC 10 20 Ripple 200 mV Housing material Stainless steel Weight approx 0 4...

Страница 46: ...ge Resolution temp reading 1 K Response time t90 10 ms for T 650 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C Temperature coefficient 0 05 K of temp reading K deviation to Tu 23 C...

Страница 47: ...d range Resolution temp reading 1 K Response time t90 10 ms for T 800 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C plus 3 0 K Temperature coefficient 0 05 K of temp reading K devi...

Страница 48: ...esponse time t90 10 ms for T 800 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C plus 3 0 K Temperature coefficient 0 05 K of temp reading K deviation to Tu 23 C Ambient temperature...

Страница 49: ...ange Resolution temp reading 1 K Response time t90 10 ms for T 800 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C plus 3 0 K Temperature coefficient 0 05 K of temp reading K deviati...

Страница 50: ...Resolution temp reading 1 K Response time t90 10 ms for T 800 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C plus 3 0 K Temperature coefficient 0 05 K of temp reading K deviation to...

Страница 51: ...Response time t90 10 ms for T 800 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C plus 3 0 K Temperature coefficient 0 05 K of temp reading K deviation to Tu 23 C Ambient temperature...

Страница 52: ...10 ms for T 700 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C plus 3 0 K Temperature coefficient 0 05 K of temp reading K deviation to Tu 23 C Ambient temperature 0 65 C Current c...

Страница 53: ...ponse time t90 10 ms for T 750 C Repeatability 3 K Measurement uncertainty 1 5 of temp reading C Temperature coefficient 0 05 K of temp reading K deviation to Tu 23 C Ambient temperature 0 65 C in sum...

Страница 54: ...e t90 10 ms for T 750 C Repeatability 3 K Measurement uncertainty 1 5 of temp reading C Temperature coefficient 0 05 K of temp reading K deviation to Tu 23 C Ambient temperature 0 65 C in sum of load...

Страница 55: ...T 750 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C Temperature coefficient 0 05 K of temp reading K deviation to Tu 23 C Ambient temperature 0 65 C in sum of load currents Do1 and...

Страница 56: ...750 C Repeatability 2 K Measurement uncertainty 1 0 of temp reading C Temperature coefficient 0 05 K of temp reading K deviation to Tu 23 C Ambient temperature 0 65 C in sum of load currents Do1 and...

Страница 57: ...quot quot quot quot d1 SP 770 C 975 C 980 C 890 C 890 C d1 rP 750 C 950 C 950 C 870 C 870 C d1 FN no no no no no d1 dS oFF oFF oFF oFF oFF d1 dr oFF oFF oFF oFF oFF d2 d2 S tu Hi tu Hi tu Hi tu Hi tu...

Страница 58: ...63 AF 1 2 PKL 68 AF 1 2 User settings s s Eps 100 100 100 100 100 s tAU 100 0 100 0 100 0 100 0 100 0 s NEN oFF oFF oFF oFF oFF s Phd 0 1 0 1 0 1 0 1 0 1 s dAp 0 0 0 0 0 0 0 0 0 0 EF diSP on on on on...

Страница 59: ...T HOLDERS AND CONTRIBUTORS AS IS AND ANY EXPRESS OR IMPLIED WARRANTIES INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PAR TICULAR PURPOSE ARE DISCLAIMED IN NO...

Страница 60: ...to any form of mechanical or electronic reproduction as well as to electronic transmission in any form through any medium Please note Unless otherwise stated in this instruction manual the instruments...