Chapter 4 Look over INS (internal statement number) of load cell
Please plug calibration plug into 15 pins RS232 at the back of indicator for load cell INS
information. The ISN provides technical parameters for manual Corner adjustment and
information for us to judge it is malfunction or not. Please refer to the below chart for
detailed information.( * is original value, load cell number starts 01, if 4 load cells in total
they are 01,02,03,04)
Step Operation Display
Remark
1. Select the
number of
load cell
Press
[check ]
Press [0][2]
Press
[check ]
[no-01]
[no-02]
Plug in calibration plug under weighing
state
Default: check cell 01 ISN
command: check cell 02 ISN
2. ISN
display
Press
[check ]
[ 2350]
No. 02 load cell 02 ISN is 02350
When finish, it returns to weighing state.
Notice
please plug off the calibration plug when operation finishes
When mistake happen which be caused by hot plug in or off, please press key “check“or
“power off” for a new set up