Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-49
delay()
This function delays the execution of the commands that follow it.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
delay(
seconds
)
seconds
The number of seconds to delay (1 to 100,000 s)
Details
You cannot set a delay for zero seconds.
The instrument delays execution of the commands for at least the specified number of seconds and fractional
seconds. However, the processing time may cause the instrument to delay 5
μs to 10 μs (typical) more than the
requested delay.
Example 1
beeper.beep(0.5, 2400)
delay(0.250)
beeper.beep(0.5, 2400)
Emit a double-beep at 2400 Hz. The sequence is
0.5 s on, 0.25 s off, 0.5 s on.
Example 2
dataqueue.clear()
dataqueue.add(35)
timer.reset()
delay(0.5)
dt = timer.measure.t()
print("Delay time was " .. dt)
print(dataqueue.next())
Clear the data queue, add 35 to it, and then delay
0.5 seconds before reading it.
Output:
Delay time was 0.500099
3.50001
Also see
None
digio.readbit()
This function reads one digital I/O line. This function is not available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
data
= digio.readbit(
N
)
data
The state of the I/O line
N
Digital I/O line number to be read (1 to 14)