1. Shut down and turn OFF your computer.

2. Install input modules, output modules, or a mixture of input and output modules in sockets K1 through K4. 

3. Set switches S1 through S4 to “IN” or “OUT” to match the type of modules installed in sockets K1 through K4. Refer to

Table 1.

 

NOTE

 

If any of sockets K1 through K4 are not used, set the corresponding switches to OUT.

 

4. Install output modules, only, in sockets K5 through K8.

5. If you have connected the SRA-01 to a PIO-12 or PIO-24 board and will be processing data via interrupts, then optionally

install an input module in socket K9 to isolate your interrupt signal source from the PIO-12/24 interrupt input.

 

WARNING

 

Install a module in socket K9 only if you have connected the SRA-01 to a PIO-12,
PIO-24, KPCI-PIO24, or KPCMCIA-PIO24 board. Otherwise, a potentially damaging
signal could be connected to pin 1 of the 37-pin D connector.

 

6. Connect your digital I/O board to the SRA-01 37-pin D connector, using available cables and accessories, as shown in

Figure 1. 

If you make your own cables, use a standard 37-pin female D connector (Keithley part number SFC-37). Refer to Figure 2
and Table 1 for pin assignments.

 

Table 1

 

Module, switch, port, and connector-pin correspondence for module sockets K1 through K9

 

Module 
socket 

Corresponding 
I/O selection 
switch 

Module type 
that is 
installed in 
socket

Required I/O 
switch setting

I/O port bit that 
is connected to 
module

Pin of 37-pin D 
connector that 
is connected to 
module

 

K1

S1

Input

IN

PC3

26

Output

OUT

PB7

3

K2

S2

Input

IN

PC2

27

Output

OUT

PB6

4

K3

S3

Input

IN

PC1

28

Output

OUT

PB5

5

K4

S4

Input

IN

PC0

29

Output

OUT

PB4

6

K5

Not applicable

Output only

Not applicable PB3

7

K6

Not applicable 

Output only

Not applicable  PB2

8

K7

Not applicable

Output only

Not applicable PB1

9

K8

Not applicable

Output only

Not applicable PB0

10

K9

Not applicable

Input only

Not applicable Interrupt input of 

PIO-12 and PIO-
24, 

 

only

 

1

Содержание SRA-01

Страница 1: ...nal source from the PIO 12 24 interrupt input LEDs monitor all module activity Safety summary WARNING Read and follow the Safety Precautions discussed at the end of this manual This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury Read the operating and safety information carefully before u...

Страница 2: ...he 37 pin D connector 6 Connect your digital I O board to the SRA 01 37 pin D connector using available cables and accessories as shown in Figure 1 If you make your own cables use a standard 37 pin female D connector Keithley part number SFC 37 Refer to Figure 2 and Table 1 for pin assignments Table 1 Module switch port and connector pin correspondence for module sockets K1 through K9 Module socke...

Страница 3: ...resent A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring Hazardous rated circuits must be provided with external double insulation or reinforced insulation Do not place accessible non hazardous I O modules next to hazardous I O modules Refer to the isolation diagram Figure 4 and to the section Safety Precautions CAB 1284CC STA 3108 D1 SRA...

Страница 4: ...HAZARDOUS Isolation A Isolated I O USER supplied wiring Isolation B Computer circuits Module at K8 NON HAZARDOUS Isolation A Isolated I O USER supplied wiring Isolation B Computer circuits Module at K9 NON HAZARDOUS Isolation A Isolated I O USER supplied wiring IsolationA per EN61010 1 1995 DOUBLE INSULATION 300V CAT II pollution degree 2 Isolation B per EN61010 1 1995 BASIC INSULATION 300V CAT II...

Страница 5: ... advanced users would be to change the 74240 chip in the SRA 01 board to a 74244 chip All output bits will be inverted and the default state will be OFF This also inverts the bits during operation and must be taken into account during program ming This is considered a non standard configuration and is not supported by Keithley In struments Sensing AC Voltage Contolling AC Voltage Use with OAC 05 m...

Страница 6: ... AWG Input modules IDC 05 IAC 05 IAC5A Nominal voltage 5 28VDC 120VAC VDC 240VAC Max input current 34mA 5mA RMS 5mA RMS Max turn ON time 1 0ms 20ms 20ms Max turn OFF time 1 0ms 30ms 30ms Output modules ODC 05 ODC5A OAC 05 OAC5A or SM OAC5A Nominal output voltage 5 48VDC 5 150VDC 120VAC 240VAC Max load current 3 0A 1 0A 3 5A RMS 3 5A RMS Max drive current 18mA 18mA 20mA 20mA Response time 100µs on ...

Страница 7: ... hazard exists when voltage levels greater than 30V RMS 42 4V peak or 60VDC are present A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring Operators of this product must be protected from electric shock at all times The responsible body must ensure that operators are prevented access and or insulated from every connection point In some cas...

Страница 8: ...t components in mains circuits including the power transformer test leads and input jacks must be purchased from Keithley Instruments Standard fuses with applicable national safety approvals may be used if the rating and type are the same Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component Note that selected pa...

Отзывы: