Introduction

 

This document provides safety information, installation instructions, and specifications for the Keithley Model SRA-01 8-Chan-
nel solid-state I/O-module accessory board. 

 

Description

 

The SRA-01 interfaces full-size, industry-standard digital I/O modules to a variety of digital I/O and combined analog/digital
I/0 boards. The digital I/O modules allow the board to sense AC and DC signals and switch AC and DC loads. The SRA-01 is
compatible with the following Keithley boards:

 

 

Digital I/O boards: KPCI-3160, KPCI-PIO96, PIO-96J, PIO-12/24, KPCI-PIO24, and KPCMCIA-PIO24

 

 

Combined analog/digital I/0 boards: KPCI-3107/3108, DAS-1600/1200, and DASCON-1

Four types of modules can be used with the SRA-01: DC input, DC output, AC input, and AC output. Sockets K1 through K4
of the SRA-01 accommodate any combination of input or output modules. (A switch for each socket configures it as either an
input-module socket or as an output-module socket.) Sockets K5 through K8 accommodate four output modules, only. If you
connect a PIO-12/24 to the SRA-01, socket K9 accommodates an additional input module to isolate an interrupt signal source
from the PIO-12/24 interrupt input. 

LEDs monitor all module activity. 

 

Safety summary

 

WARNING

 

Read and follow the “Safety Precautions” discussed at the end of this manual.

 

This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read the operating and safety information carefully before using the product.

Users of this product must be protected from electric shock at all times. The responsible body must ensure that users are pre-
vented access and/or insulated from every connection point.

Do not connect switching cards directly to unlimited power circuits. When connecting sources to switching cards, install pro-
tective devices to limit fault current and voltage to the card.

 

Installation

 

WARNING

 

Users of this product must be protected from electric shock at all times. The responsible
body must ensure that users are prevented access and/or insulated from every connection
point. In some cases, connections must be exposed to potential human contact. Product
users in these circumstances must be trained to protect themselves from the risk of elec-
tric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part
of the circuit may be exposed.

 

CAUTION

 

Ensure that the computer power is turned OFF before installing the SRA-01. Connect-
ing the SRA-01 to the computer while the power is ON can damage your computer, the
accessory, or both.

 

PA-639 Rev. F / 6-02

Содержание SRA-01

Страница 1: ...nal source from the PIO 12 24 interrupt input LEDs monitor all module activity Safety summary WARNING Read and follow the Safety Precautions discussed at the end of this manual This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury Read the operating and safety information carefully before u...

Страница 2: ...he 37 pin D connector 6 Connect your digital I O board to the SRA 01 37 pin D connector using available cables and accessories as shown in Figure 1 If you make your own cables use a standard 37 pin female D connector Keithley part number SFC 37 Refer to Figure 2 and Table 1 for pin assignments Table 1 Module switch port and connector pin correspondence for module sockets K1 through K9 Module socke...

Страница 3: ...resent A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring Hazardous rated circuits must be provided with external double insulation or reinforced insulation Do not place accessible non hazardous I O modules next to hazardous I O modules Refer to the isolation diagram Figure 4 and to the section Safety Precautions CAB 1284CC STA 3108 D1 SRA...

Страница 4: ...HAZARDOUS Isolation A Isolated I O USER supplied wiring Isolation B Computer circuits Module at K8 NON HAZARDOUS Isolation A Isolated I O USER supplied wiring Isolation B Computer circuits Module at K9 NON HAZARDOUS Isolation A Isolated I O USER supplied wiring IsolationA per EN61010 1 1995 DOUBLE INSULATION 300V CAT II pollution degree 2 Isolation B per EN61010 1 1995 BASIC INSULATION 300V CAT II...

Страница 5: ... advanced users would be to change the 74240 chip in the SRA 01 board to a 74244 chip All output bits will be inverted and the default state will be OFF This also inverts the bits during operation and must be taken into account during program ming This is considered a non standard configuration and is not supported by Keithley In struments Sensing AC Voltage Contolling AC Voltage Use with OAC 05 m...

Страница 6: ... AWG Input modules IDC 05 IAC 05 IAC5A Nominal voltage 5 28VDC 120VAC VDC 240VAC Max input current 34mA 5mA RMS 5mA RMS Max turn ON time 1 0ms 20ms 20ms Max turn OFF time 1 0ms 30ms 30ms Output modules ODC 05 ODC5A OAC 05 OAC5A or SM OAC5A Nominal output voltage 5 48VDC 5 150VDC 120VAC 240VAC Max load current 3 0A 1 0A 3 5A RMS 3 5A RMS Max drive current 18mA 18mA 20mA 20mA Response time 100µs on ...

Страница 7: ... hazard exists when voltage levels greater than 30V RMS 42 4V peak or 60VDC are present A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring Operators of this product must be protected from electric shock at all times The responsible body must ensure that operators are prevented access and or insulated from every connection point In some cas...

Страница 8: ...t components in mains circuits including the power transformer test leads and input jacks must be purchased from Keithley Instruments Standard fuses with applicable national safety approvals may be used if the rating and type are the same Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component Note that selected pa...