Calibration 2-5
4.
Once the correct password is displayed, press the ENTER key. If the password was cor-
rectly entered, the following message will be displayed:
CALIBRATION UNLOCKED
Calibration can now be executed.
5.
Press EXIT to return to normal display. Calibration will be unlocked and assume the
states summarized in Table 2-2. Attempts to change any of the settings listed below with
calibration unlocked will result in an error +510, "Not permitted with cal un-locked."
NOTE
With calibration unlocked, the sense function and range track the source function and
range. That is, when :SOUR:FUNC is set to VOLT, the :SENS:FUNC setting will be
'VOLT:DC'. When :SOUR:FUNC is set to CURR, the :SENS:FUNC setting will be
'CURR:DC'. A similar command coupling exists for :SOUR:VOLT:RANG/
:SENS:VOLT:RANG and SOUR:CURR:RANG:SENS:CURR:RANG.
Unlocking calibration by remote
To unlock calibration via remote, send the following command:
:CAL:PROT:CODE '<password>'
For example, the following command uses the default password:
:CAL:PROT:CODE 'KI002420'
Table 2-2
Calibration unlocked states
Mode
State
Equivalent remote command
Concurrent Functions
Sense Function
Sense Volts NPLC
Sense Volts Range
Sense Current NPLC
Sense Current Range
Filter Count
Filter Control
Filter Averaging
Source V Mode
Volts Autorange
Source I Mode
Current Autorange
Autozero
Trigger Arm Count
Trigger Arm Source
Trigger Count
Trigger Source
OFF
Source
1.0
Source V
1.0
Source I
10
REPEAT
ON
FIXED
OFF
FIXED
OFF
ON
1
Immediate
1
Immediate
:SENS:FUNC:CONC OFF
:SENS:FUNC <source_ function>
:SENS:VOLT:NPLC 1.0
:SENS:VOLT:RANG <source_V_range>
:SENS:CURR:NPLC 1.0
:SENS:CURR:RANG <source_ I_ range>
:SENS:AVER:COUN 10
:SENS:AVER:TCON REPeat
:SENS:AVER:STAT ON
:SOUR:VOLT:MODE FIXED
:SOUR:VOLT:RANG:AUTO OFF
:SOUR:CURR:MODE FIXED
:SOUR:CURR:RANG:AUTO OFF
:SYST:AZERO ON
:ARM:COUNT 1
:ARM:SOUR IMMediate
:TRIG:COUNT 1
:TRIG:SOUR IMMediate
Содержание SourceMeter 3A 2420
Страница 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 15: ...1 Performance Verification...
Страница 31: ...2 Calibration...
Страница 53: ...3 Routine Maintenance...
Страница 56: ...3 4 Routine Maintenance...
Страница 57: ...4 Troubleshooting...
Страница 70: ...4 14 Troubleshooting...
Страница 71: ...5 Disassembly...
Страница 78: ...5 8 Disassembly...
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ...6 Replaceable Parts...
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ...A Specifications...
Страница 112: ...A 6 Specifications...
Страница 113: ...B Command Reference...
Страница 124: ...B 12 Command Reference...
Страница 125: ...C Calibration Programs...