Safety

Introduction

Models 2634B, 2635B, and 2636B

System SourceMeter Instruments

Quick Start Guide

Содержание SourceMeter 2635B

Страница 1: ...Safety Introduction Models 2634B 2635B and 2636B System SourceMeter Instruments Quick Start Guide...

Страница 2: ...require protection for high transient overvoltages often associated with localAC mains connections Certain Keithley measuring instruments may be connected to mains These instruments will be marked as...

Страница 3: ...omes contaminated and operation is affected the board should be returned to the factory for proper cleaning servicing Safety precaution revision as of March 2016 Do not touch any object that could pro...

Страница 4: ...liance levels may result in damage to the instrument or to the device under test Power and environmental specifications For indoor use only Power Supply 100 VAC to 240 VAC 50 Hz to 60 Hz auto sensing...

Страница 5: ...units SMUs that simplify test processes by combining source and measure capabilities in a single instrument A Series 2600B SourceMeter Instrument is a scalable high throughput cost effective solution...

Страница 6: ...he box for damage 2 Open the top of the box 3 Remove the documentation and accessories 4 Carefully lift the instrument out of the box 5 Remove the packaging insert 6 Inspect the instrument for any obv...

Страница 7: ...the 2634B and 2635B 4 Power line cord 5 One triaxial to alligator clip cable for the 3635B Two triaxial to alligator clip cables for the 2634B and 2636B 6 16 AWG black grounding wire 7 3 pin screw ter...

Страница 8: ...e in an unknown state until they are properly initialized Make sure the design can tolerate this situation without causing operator injury or hardware damage Connect the instrument Important test syst...

Страница 9: ...o set Make sure the operating voltage in your area is compatible To keep users safe always read and follow all safety warnings provided with each of the instruments in your system The power cord suppl...

Страница 10: ...ord to the power connection on the rear panel 3 Connect the plug of the power cord to a grounded AC outlet Operating the instrument on an incorrect line voltage may cause damage to the instrument poss...

Страница 11: ...the display will show KEITHLEY Model 2635B Line frequency detection and other startup checks The entire power up process takes approximately 30 seconds to complete When initialization is complete if t...

Страница 12: ...T for this test Step 1 Set source function range and level 1 Press the SRC key You will see a blinking character in the SrcA value field Confirm that mV is displayed if not press the SRC key again 2 W...

Страница 13: ...el to set the source value to 20 0000 V and then press the navigation wheel to enter the selection and exit EDIT mode Step 2 Set the source limit 1 Press the LIMIT key You will see a blinking characte...

Страница 14: ...ected the Series 2600B automatically selects the best range for the measured value You will briefly see the display shown below and then the main display screen reappears Step 4 Turn the output on Tur...

Страница 15: ...r LED turns off For a basic understanding of the of the Series 2600B functionality we strongly recommend that first time users become familiar with the Series 2600B Reference Manual These steps confir...

Страница 16: ...ont panel The included USB flash drive contains a copy of the 2400 personality script which allows the Series 2600B instrument to accept 2400 SCPI commands For more information about loading and runni...

Страница 17: ...to the Series 2600B Reference Manual for detailed information about all features of the instrument Keithley Instruments website See www tek com keithley for support and additional information about t...

Страница 18: ...00800 2255 4835 New Zealand 0800 800 238 Norway 800 16098 People s Republic of China 400 820 5835 Philippines 1 800 1601 0077 Poland 41 52 675 3777 Portugal 80 08 12370 Republic of Korea 82 2 6917 50...