*See full specifications for other pulse regions.
Carefully consider and configure the appropriate
output-off state, source levels, and compliance
levels before connecting the instrument to
a device that can deliver energy. Failure to
consider the output-off state, source levels, and
compliance levels may result in damage to the
instrument or to the device under test.
Carefully consider and configure the appropriate
output-off state, source levels, and compliance
levels before connecting the instrument to
a device that can deliver energy. Failure to
consider the output-off state, source levels, and
compliance levels may result in damage to the
instrument or to the device under test.
Power and environmental specifications
For indoor use only.
Power Supply
100 VAC to 240 VAC, 50 Hz to 60 Hz
(auto sensing)
Maximum VA
240 VA
Operating altitude
Maximum 2000 m (6562 feet) above
sea level
Operating
temperature
0 °C to 50 °C (32 °F to 122 °F),
70% relative humidity up to 35 °C.
Derate 3% relative humidity,
35 °C to 50 °C
Storage temperature
25 °C to 65 °C (77 °F to 149 °F)
Pollution degree
1 or 2
DC source output
30.3 W maximum
±200 VDC maximum
±1.515 A at ±20 VDC
±101 mA at ±200 VDC
Pulse region 4*
Region maximums: 10 A at 5 V
Maximum pulse width: 1 ms
Maximum duty cycle: 2.2%
Measurement input
Measurement category I
Voltage: 200 VDC maximum HI to LO,
250 VDC LO to ground
Current: 101 mA at 200 V;
1.515 A at 20 V
Impedance: Variable