13-4 Factory Scripts
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
TSP Project Name:
KIFactoryGeneral
TSP Test Script Name:
KIGeneral
Firmware Version:
1.0.2 and later
Function:
PulseVMeasureI(smu, bias, level, ton, toff, points)
Description: Performs a specified number of pulse V, measure I cycles:
• Sets the
smu
to output
bias
volts and dwell for
ton
seconds.
• Sets the
smu
to output
level
volts and dwell for
ton
seconds.
• Performs current measurement with source at
level
amps.
• Sets the
smu
to output
bias
volts for
toff
seconds.
• Repeats the above sequence for
points
pulse-measure cycles.
level
bias
ton
bias
ton
toff
Current measurement
Parameters:
smu, bias, level, ton, toff, points
smu:
SourceMeter Channel (A or B). Defaults to smua if all parameters
are omitted when function is called.
bias:
Bias level in volts.
level:
Pulse level in volts.
ton:
Pulse on-time in seconds.
toff:
Pulse off-time in seconds.
points:
Number of pulse-measure cycles.
Data: Pulsed current measurements, voltage levels and timestamps are stored in
smuX.nvbuffer1
.
Example:
PulseVMeasureI(smub, -1, 1, 1E-3, 2E-3, 20)
SMU B will output -1V and dwell for 1ms, output 1V and dwell for 2ms, and
then perform a current measurement. After the measurement, the output will
return to -1V and dwell for 2ms. This pulse-measure process will repeat 19
more times.