12-48 Instrument Control Library
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
localnode.linefreqauto
Attribute
Automatic power line frequency detection control.
Usage
flag = localnode.linefreqauto
-- Read auto line frequency detection
-- setting.
localnode.linefreqauto = true
-- Enable automatic line frequency
-- detection at powerup.
Remarks
• When this attribute is set to true, the power line frequency will be detected
automatically the next time the SourceMeter powers up.
• After the power line frequency is automatically detected at power-up, the
localnode.linefreq attribute
will be set automatically to 50 or 60.
• If the
localnode.linefreq attribute
is explicitly set,
localnode.linefreqauto
will be automatically set to false.
• When used in an expanded system (TSP-Link),
localnode.linefreqauto
is
sent to the Remote Master node only.
• Use
node[N].linefreqauto
(where N is the node number) to
send the command to any node in the system. See
for details on TSPLink.
• This command is only available in firmware version 1.2.0 and later.
Also see
Example
Disable automatic power line frequency detection:
localnode.linefreqauto = 0