
Model 2460 Interactive SourceMeter® Instrument User's Manual
Section 8: Measuring I-V characteristics of a solar panel
2460-900-01 Rev. A / September 2014
8-5
7. Press the
MENU
key.
8. Under Measure, select
Settings
.
9. Select the button next to Sense Mode and select
4-Wire Sense
.
10. Press the
MENU
key.
11. Under Source, select
Sweep
.
12. Set the Start level to
0 V
and select
OK
.
13. Set the Stop level to
20 V
and select
OK
.
14. Select the button next to Definition and select
Number of Points
.
15. Select the button next to Points, enter
115
, and select
OK
.
16. Swipe the SWEEP SETTINGS screen until you see Source Limit.
17. Select the button next to Source Limit, enter
4
A
, and select
OK
.
18. Select the button next to Delay and select
Specify Delay Manually
.
19. Enter
50 ms
and select
OK
.
20. Select
Generate
. This sets up a trigger model for the sweep.
21. Press the
MENU
key.
22. Under Measure, select
Data Buffers
.
23. Select
defbuffer1
to open the Settings for defbuffer1 dialog box.
24. Select
Clear Buffer,
select
Yes
, and select
OK
.
25. Press the
MENU
key.
26. Under Views, select
Graph
.
27. Press the
TRIGGER
key to initiate the trigger model. The output turns on and a RUN indicator is
visible at the top of the screen while the sweep is running.
28. Press the trigger key again to repeat the sweep.
The following figure shows an example of solar panel I-V measurements on the front-panel graph.
Note that the current is negative on the graph because the Model 2460 is sinking current.
Figure 45: Example of solar panel measurements on the Model 2460 front-panel graph