Model 2460 Interactive SourceMeter® Instrument User's Manual
Section 3: Using a remote interface
2460-900-01 Rev. A / September 2014
3-13
LAN troubleshooting suggestions
If you are unable to connect to the web interface of the instrument, check the following items:
•
Verify that the network cable is in the LAN port on the rear panel of the instrument, not one of the
TSP-Link
®
ports.
•
Verify that the network cable is in the correct port on the computer. The LAN port of a laptop may
be disabled when the laptop is in a docking station.
•
Verify that the setup procedure used the configuration information for the correct ethernet card.
•
Verify that the network card of the computer is enabled.
•
Verify that the IP address of the instrument is compatible with the IP address on the computer.
•
Verify that the subnet mask address of the instrument is the same as the subnet mask address of
the computer.
•
Restart your computer.
•
Turn the instrument's power off, and then on. Wait at least 60 seconds for the network
configuration to be completed. Verify that the correct settings are assigned to the instrument:
1. Press the
MENU
key.
2. Under System, select
Communication
.
3. Select the
LAN
tab.
4. Verify the settings.
If the above actions do not correct the problem, contact your system administrator.