1-24
Return to
Section Topics
2400S-903-01 Rev. E / September 2011
Section 1: Introduction
Series 2400 SourceMeter® Quick Start Guide
Figure 1-8
Logarithmic staircase sweep (example)
Custom sweep
For a custom sweep, the user specifies the number of point in a sweep, and the source level at
each point. Figure 9 shows a 6-point, 50% duty cycle pulse sweep. The specified voltage level at
points P0, P2 and P4 is 1V, while the voltage level at P1, P3 and P5 is 0V.
With the desired source (V or I) selected, use the following menu sequence to specify the number
of sweep points and the source level at each point:
Press CONFIG > press SWEEP > select TYPE > select CUSTOM
After selecting CUSTOM, these menu items will be displayed:
# POINTS - Use to specify the number of points in the sweep (1 to 2500)
ADJUST POINTS - Use to set the source level at each point in the sweep. The first point in the
sweep is P0000. After setting the level for P0000, increment the pointer to P0001 and set the level
for that point. In a similar manner, set the level for each point in the sweep.
INIT - This option allows you to set a consecutive range of sweep points to a specified level. After
selecting INIT, you will be prompted to enter the source VALUE, the START PT (point), and the
STOP PT (point). For example, if you want points 10 through 15 to be 1V, set VALUE to 1V, START
PT to 10, and STOP PT to 15.
Delay
Delay
Delay
Delay
Start
Bias
Stop
(10)
Measure
#1
Measure
#2
Measure
#3
Measure
#5
X = Measurement Point
X
X
X
X
Delay
X
Measure
#4
Log Points = 5
1
10
5.6234
3.1623
1.7783
Volts
Log
Scale