Operating Basics
NOTE.
All measurements are at the terminals. If the wires used to connect the channels together are too small, too long,
or improperly tightened, accuracy of measurements will be adversely affected.
1.
Wire the outputs in parallel. (See page 24,
Wiring in Parallel
.)
2.
Push
Menu
and use the up arrow key to navigate to
Combine CH1+CH2
.
3.
Push
Enter
.
4.
Use the down arrow key to navigate to
I1+I2 Parallel
and push
Enter
.
5.
Check that the word
Para
appears on the display, replacing the Channel 2 voltage and current readings. This indicates
that the power supply is in the I1 + I2 state. The total output current is displayed on Channel 1.
6.
Set the Channel 1 current to the desired total current value (up to 3 A).
User Settings
Recall the Power Supply ON/OFF Output State at Power On
This parameter determines the On or Off output state after the power supply is powered on. If you select
On
, the power
supply will restore the state of the output to that which was in use before the power was last turned off. If the output was
turned
On
when the power supply is turned off or loses power then the output will return to the
On
state when the power
supply is turned back on or power is restored.
Off
will disable this function and the output channel will always be set
to
Off
after the power supply is powered on.
To enable or disable this control,
1.
Push
Menu
.
2.
Use the up and down arrow keys to select
>User Settings
.
3.
Push
Enter
.
4.
Use the up and down arrow keys to select
Output Recall
.
5.
Push
Enter
.
6.
Use the up or down arrow keys to select
On
or
Off
.
7.
Push
Enter
.
8.
Push
Esc
to exit the menu system.
NOTE.
The default selection is set to Off.
Recall the Power Supply Operating Parameters at Power On
This parameter determines whether the power supply saves its most recent settings, such as voltage and current, and
restores these settings at power on. If you set this parameter to
Off
then the power supply returns to the default settings
at power on. If you select
On
, the power supply will restore the state to that which was in use before the power was
last turned off.
Series 2200 Multi-Channel Programmable DC Power Supplies User Manual