background image

Document number: S46-901-01 Rev. F / March 2011

4-5

S46/S46T/S46L Microwave Switch System Instruction Manual

Section 4: Service Information

NOTE

Make sure that you order the relay kit for the applicable system (S46, S46T, or S46L). 
Relay kits contain components and instructions appropriate for your specific system.

Relay replacement

WARNING

Relay replacement should be performed by qualified service personnel 
only.  Do not perform this procedure unless you are qualified to do so.

Before replacing a relay, turn off the power and disconnect the line cord 
from the instrument. Remove any signal connections from all relays

Table 4-16
S46/S46T/S46L relay kits

System

Relay kit model number

Description

Relay location

S46

S46-SPDT-KIT

18GHz SPDT

1 to 8 

S46-SP4T-KIT

18GHz SP4T

A to D 

S46-SP6T-KIT

18GHz SP6T

A to D 

S46-SPDT-KIT-R 

18GHz SPDT

1 to 8 

S46-SP4T-KIT-R  

18GHz SP4T

A to D 

S46-SP6T-KIT-R  

18GHz SP6T

A to D 

S46-SPDT-KIT-26 

26.5GHz SPDT

1 to 8 

S46-SP4T-KIT-26 

26.5GHz SP4T

A to D 

S46-SP6T-KIT-26 

26.5GHz SP6T

A to D 

S46-SPDT-KIT-40

40GHz SPDT

1 to 8 

S46-SP4T-KIT-40

40GHz SP4T

A to D 

S46-SP6T-KIT-40

40GHz SP6T

A to D 

S46T

S46T-SPDT-KIT

18GHz SPDT

1 to 8 

S46T-SP4T-KIT

18GHz SP4T

A to D 

S46T-SP6T-KIT

18GHz SP6T

A to D 

S46T-SPDT-KIT-26

26.5GHz SPDT

1 to 8 

S46T-SP4T-KIT-26

26.5GHz SP4T

A to D 

S46T-SP6T-KIT-26

26.5GHz SP6T

A to D 

S46T-SPDT-KIT-T

18GHz SPDT

1 to 8 

S46T-SPDT-KIT-26T

26.5GHz SPDT

1 to 8  

S46T-SP4T-KIT-T

18GHz SP4T

A to D

S46T-SP6T-KIT-T

18GHz SP6T

A to D

S46T-SP4T-KIT-26T

26.5GHz SP4T

A to D

S46T-SP6T-KIT-26T

26.5GHz SP6T

A to D

S46T-XFER-KIT 

18GHz XFER

A to D

S46T-XFER-KIT-26 

26.5GHz XFER

A to D

S46T-MSPDT-KIT 

18GHz SPDT

A to D

S46T-MSPDT-KIT-26

26.5GHz SPDT

A to D

S46L

S46L-SPDT-KIT-L

20 Ghz SPDT

1 to 4

S46L-SP6T-KIT-T

18 Ghz SP6T

A to D

Содержание S46

Страница 1: ...www keithley com A G R E A T E R M E A S U R E O F C O N F I D E N C E S46 S46T S46L Microwave Switch System Instruction Manual S46 901 01 Rev F March 2011...

Страница 2: ...ts Inc All rights reserved Cleveland Ohio U S A Document number S46 901 01 Rev F March 2011 All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments Inc O...

Страница 3: ...ith electrical signals that are rated Measurement Category I and Measurement Category II as described in the International Electrotechnical Commission IEC Standard IEC 60664 Most measurement control a...

Страница 4: ...k of electric shock Use standard safety precautions to avoid personal contact with these voltages The symbol on an instrument shows that the surface may be hot Avoid personal contact to prevent burns...

Страница 5: ...1 Handling precautions 2 1 System configuration 2 1 Layout 2 1 Simplified schematic 2 3 Connections 2 4 Signal connections 2 6 GPIB control connection 2 6 Power connections 2 8 Ground connection 2 11...

Страница 6: ...s 3 20 Common commands 3 20 Errors 3 26 4 Service Information 4 1 Introduction 4 1 Handling and cleaning precautions 4 1 Handling precautions 4 1 Circuit board and connector cleaning 4 1 Coaxial switc...

Страница 7: ...nated coaxial microwave relays and four multi pole unterminated coaxial microwave relays Any of these multi pole unterminated relays can be one of the following relay types SP4T or SP6T S46 switching...

Страница 8: ...shipment Shipment contents The following items are included with every S46 S46T S46L order S46 S46T S46L Microwave Switch System includes power cord Product Information CD ROM part number S46 950 01 w...

Страница 9: ...KIT Standard performance 18 GHz unterminated SPDT relay and control cable assembly S46 SP4T KIT Standard performance 18 GHz unterminated SP4T relay and control cable assembly S46 SP6T KIT Standard pe...

Страница 10: ...relays mounting plate and control cable assembly Note Kit applicable only for relay A D mounting locations S46T SP4T KIT 18 GHz unterminated SP4T relay mounting plate and control cable assembly S46T S...

Страница 11: ...sure the S46 S46T S46L is turned off the line cord is disconnected and it is not connected to any external circuitry The power supply cord must be used as the mains disconnecting device The plug shoul...

Страница 12: ...nt kit CAUTION Installation of the rear brackets is recommended to properly support the S46 S46T S46L Foot mounting To operate the S46 S46T S46L on a bench attach the supplied rubber feet to the botto...

Страница 13: ...e careful when handling the S46 S46T S46L to avoid contamination from foreign materials such as body oils Such contamination can reduce isolation resistance To avoid possible contamination Do not touc...

Страница 14: ...pole relay connections SP4T SP3T SP5T SP6T S46 MICROWAVE SWITCH SYSTEM PWR 1 3 5 1 3 5 1 3 5 1 3 5 R1 R3 R5 R7 ERR 2 4 6 2 4 6 2 4 6 2 4 6 R2 R4 R6 R8 RA RB RC RD RELAY 1 RELAY 2 RELAY 3 RELAY 4 NO C...

Страница 15: ...and one common RELAY A through RELAY D on the S46 and S46T can be 4 pole 6 pole or MSPDT relays each with a maximum of six inputs and one common The number of inputs depends on relay configuration as...

Страница 16: ...6 5 GHz and unterminated 40 GHz relays have female SMA 2 9 connectors All others have female SMA connectors RELAY A to RELAY D 4 pole or 6 pole RELAY 1 to RELAY 8 2 pole Normally Closed Normally Open...

Страница 17: ...RELAY 1 through RELAY 4 for S46L are populated with 2 pole RF relays Any or all relay locations can be populated Channel assignments Table 2 1 summarizes relay channel assignments See Section 3 for in...

Страница 18: ...IB control connection The rear panel IEEE 488 GPIB control port is connected to the GPIB port of a computer controller using a shielded IEEE 488 interface cable with metric mating screws Figure 2 5 sh...

Страница 19: ...und is 0 V Connectors may be stacked to allow a number of parallel connections to one instrument Two screws located on a standard connector maintain secure connections between connectors NOTE To minim...

Страница 20: ...ion is automatic CAUTION Operating the S46 S46T S46L on an incorrect line voltage may cause damage possibly voiding the warranty Table 2 2 GPIB control connector terminals Contact number IEEE 488 desi...

Страница 21: ...before changing the line fuse To replace the line fuse 1 The fuse is located in a holder in the power module adjacent to the AC receptacle see Figure 2 2 At the right of the fuse holder is a small tab...

Страница 22: ...der see Figure 2 6 3 Twist the fuse holder to open it and then remove the fuse 4 Replace the damaged fuse with one that has the fuse rating listed in Table 2 4 5 Replace the top cover WARNING For cont...

Страница 23: ...iously affect their integrity The following paragraphs discuss these effects and ways to minimize them CAUTION Do not close more than one RF path per multiport switch Degradation of RF performance wil...

Страница 24: ...ource resistance and path isolation resistance Also leakage currents can be generated through these resistances by voltage sources in the system Refer to the specifications located on the Keithley Ins...

Страница 25: ...cations of the switch Maximum signal level specifications Maximum voltage 30 V DC 42 V peak Maximum Power per RF channel 35 W CW at 18 GHz 1 W switching load VSWR of 1 20 1 maximum 10 W CW at 26 5 GHz...

Страница 26: ...ration NOTE Momentarily remove power from the instrument after changing the address to update the address to the new setting Figure 3 7 Primary address DIP switch example Programming syntax Syntax rul...

Страница 27: ...ions require several command words Single command message This program message uses the command words required to perform a single programming operation Example SYST ERR Reads the system error queue M...

Страница 28: ...physical connections Figure 3 8 Command diagram In this example commands to close channels 2 and 8 are sent which closes inputs to both RELAY A and RELAY B Program message terminator PMT Each program...

Страница 29: ...multiple command program message are executed Valid commands following an invalid command in a multiple command program message are ignored Response messages A response message is the message sent by...

Страница 30: ...the instrument to the computer 1 Send the appropriate query command s in a program message 2 Address the S46 S46T S46L to talk Rule 2 The computer must receive the complete response message before ano...

Страница 31: ...Description This command query only is used to return 32 28 for the S46L comma separated values representing the closure count the number of times the channel has been closed Unpopulated channels alwa...

Страница 32: ...in the system Description This command is used to define which relays are present in the system as a clist This command need be used only if the system is being configured differently from the factor...

Страница 33: ...nding SPARameter You can determine whether or not a relay is populated by using the CONFigure CPOLe query If a relay is replaced or added after the S46L has left the factory you must update the serial...

Страница 34: ...rst in first out register Every time the error queue is queried the oldest message is read and removed from the queue If the error queue becomes full the message 350 Queue Overflow will occupy the las...

Страница 35: ...other error messages are enabled and will go into the error queue as they occur unless disabled For example to disable the 110 140 and 222 messages send STAT QUE DIS 110 140 222 Query This query comma...

Страница 36: ...ueue is empty If empty the message 0 No Error is placed in the error queue The messages in the error queue are preceded by a number Refer to Table 3 14 for a listing of error numbers and messages VERs...

Страница 37: ...ster will set to 1 which in turn sets the summary bit of the Status Byte Register 0 Output Queue Error Queue Status Byte Register Service Request Enable Register 1 EAV MAV 3 ESB RQS MSS 7 STB 0 1 EAV...

Страница 38: ...n commands to program and read the standard event register Queues The relay unit uses an output queue and an error queue The response messages to query commands are placed in the output queue As vario...

Страница 39: ...llowing SCPI query commands and then addressing the S46 S46T S46L to talk SYSTem ERRor STATus QUEue Messages in the error queue are stored in a FIFO First In First Out manner The commands to read the...

Страница 40: ...these registers Table 3 10 SCPI commands Error queue Command Description Default STATus QUEue NEXT ENABle lis t ENABle DIS able list DISable CLEar STATus subsystem Read error queue Read and clear olde...

Страница 41: ...tain the status byte also known as the serial poll byte B6 is the RQS bit See Serial polling and SRQ for details about using the serial poll sequence When using the STB command see Figure 3 11 to read...

Страница 42: ...SRQ service request In your test program you can periodically read the status byte to check if an SRQ has occurred and what caused it If an SRQ occurs the program can for example branch to an appropri...

Страница 43: ...r way to determine the decimal value is to add up the decimal weights for the bits that you wish to set Note that Table 3 12 includes the decimal weight for each register bit To set bits B5 B3 and B2...

Страница 44: ...8 211 1024 210 512 29 256 28 Table 3 13 IEEE Std 488 2 common commands and queries Mnemonic Name Description CLS Clear status Clear the Standard Event Register and error queue ESE NRf Event enable com...

Страница 45: ...andard event is unmasked enabled the occurrence of that event sets the ESB bit For information about the Standard Event Register and descriptions of the standard event bits see ESR Event Status Regist...

Страница 46: ...10000 From this binary value bits B4 and B5 of the Standard Event Status Register are set These bits indicate that a device dependent error and command error have occurred The bits of the Standard Eve...

Страница 47: ...s register has been read IDN identification query Reads identification code The identification code includes the manufacturer model number serial number and firmware revision levels Identification cod...

Страница 48: ...t Enable Register set an SRQ occurs when the corresponding bit in the Status Byte Register is set by an appropriate event For more information about register structure see the information presented ea...

Страница 49: ...t used Bit 2 error available EAV A set bit indicates that an error or status message is present in the error queue The message can be read using one of the following SCPI commands Bit 3 Not used Bit 4...

Страница 50: ...ecution of subsequent commands while device operations of the overlapped command are still in progress The WAI command is used to suspend the execution of subsequent commands until the device operatio...

Страница 51: ...cter data error EE 128 Numeric data not allowed EE 124 Too many digits EE 123 Exponent too large EE 121 Invalid character in number EE 120 Numeric data error EE 113 Undefined header EE 112 Program mne...

Страница 52: ...cautions should be observed when servicing the S46 S46T S46L Handling precautions Observe the following precautions when handling the S46 S46T S46L Do not touch connector insulators Do not touch any b...

Страница 53: ...rborne deposits clean with a cotton swab dipped in clean methanol NOTE Failure of any performance verification test may indicate that the instrument is contaminated See Handling and cleaning precautio...

Страница 54: ...a null value Leave REL enabled for the entire test 7 Remove the SMA F F adapter 8 Connect the cables to the switch common port and port under test 9 Close the switch port 10 Note the resistance readin...

Страница 55: ...3A 250V AMPS HI LO INPUTS 350V PEAK 1000V PEAK SHIFT LOCAL POWER RANGE R SHIFT CH1 REM TALK LSTN SRQ STAT REL FILT 4W BUFFER MATH REAR SCAN TIMER STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10 HOLD TRIG F...

Страница 56: ...T KIT 18GHz SP6T A to D S46 SPDT KIT R 18GHz SPDT 1 to 8 S46 SP4T KIT R 18GHz SP4T A to D S46 SP6T KIT R 18GHz SP6T A to D S46 SPDT KIT 26 26 5GHz SPDT 1 to 8 S46 SP4T KIT 26 26 5GHz SP4T A to D S46 S...

Страница 57: ...ore attempting to remove a relay use an appropriate desoldering tool to clear each mounting hole completely free of solder Soldering considerations When using solder on the circuit board observe the f...

Страница 58: ...ector nuts from the rear panel 6 Remove the digital board from the chassis 7 Remove the screws that secure the power supply to the chassis then remove the power supply To assemble the S46 S46T S46L re...

Страница 59: ...1 Section 4 Service Information S46 S46T S46L Microwave Switch System Instruction Manual Figure 4 19 S46T exploded view Top cover Relays Relays NOTE Refer to Figure 4 2 for exploded views of Digital b...

Страница 60: ...S46L Serial number Part description Component designation if applicable Keithley part number Factory service If you need to return your S46 S46T S46L to Keithley Instruments for repair perform the fo...

Страница 61: ...egisters 3 14 Environmental conditions 4 2 ERRor 3 12 Error queue 3 15 Event register sets 3 13 Example command 3 4 Exploded view 4 7 4 8 4 9 F Factory service 4 9 Flux 4 6 G General information 1 1 G...

Страница 62: ...onents 4 4 Response message terminator RMT 3 6 Response messages 3 5 RFI EMI 2 12 RS 232 connector terminals 2 8 S Schematic 2 3 Sending a response message 3 5 Serial polling and SRQ 3 18 Service requ...

Страница 63: ...of Keithley Instruments Inc All other trademarks and trade names are the property of their respective companies A G R E A T E R M E A S U R E O F C O N F I D E N C E Keithley Instruments Inc Corporate...

Отзывы: