Wiring Signals
37
4
4
4
4
4
4
4
4
Connecting Differential Voltage Inputs
shows how to connect differential voltage inputs (channels
0 and 1, in this case) to the BNC connectors on a KUSB-3116 module.
Figure 10: Connecting Differential Inputs to the BNC Connectors
shows how to connect a floating signal source to your
own screw terminal panel using differential inputs. (A floating signal
source is a voltage source that has no connection with earth ground.)
Note:
For floating signal sources, it is recommended that you
provide a bias return path for the differential channels by using the
Open Layers Control Panel applet to enable 10 k
Ω
of termination
resistance. For more information, refer to
Signal
Source
Analog In 0
Analog In 1
AD0
Analog Input
AD1
AD
2
AD
6
AD
3
AD5
AD7
AD
4
Note that the BNC connection box
automatically connects the Analog
Ground and Analog Input Return signals
appropriately.
KUSB-3116 Module
Содержание KUSB-3116
Страница 10: ...Contents x...
Страница 14: ...About this Manual xiv...
Страница 15: ...1 1 Overview Hardware Features 2 Supported Software 4 Getting Started Procedure 5...
Страница 20: ...Chapter 1 6...
Страница 21: ...7 2 Preparing to Use a Module Unpacking 9 Checking the System Requirements 10 Installing the Software 11...
Страница 26: ...Chapter 2 12...
Страница 64: ...Chapter 4 50...
Страница 65: ...51 5 Verifying the Operation of a Module Overview 53 Running the Quick Data Acq Application 54...
Страница 75: ...61 A Ground Power and Isolation Connections...
Страница 78: ...Appendix A 64...
Страница 82: ...Index 68...