KPCI-3101 — KPCI-3104 Series User’s Manual
10. Measure and compare the analog output voltages as indicated in
:
a. Measure the voltages at analog outputs 0 and 1 with your DVM/DMM.
b. Compare the voltages you measured in step 10a with the voltages you set via the analog-
output level control.
11. Using your mouse, slide the CH0 and CH1 level control buttons until the tiny displays at the
bottoms of the level controls read 5.0.
12. Measure and compare the analog output voltages as indicated in
:
a. Measure the voltages at analog outputs 0 and 1 with your DVM/DMM.
b. Compare the voltages you measured in step 12a with the voltages you set via the analog-
output level control.
13. Based on the measured voltages in steps 10 and 12, take action as follows:
•
If the voltages measured with the DVM/DMM do not agree with the level control set-
tings, then there is an apparent problem with the analog output part of your board. Stop
here, and return to the problem isolation step in “
Problem isolation Scheme F: the board
that asked you to perform the Analog output hardware test.
•
If the voltages measured with the DVM/DMM agree with level control settings, then the
analog outputs are working satisfactorily. Stop here, and return to the problem isolation
step in “
Problem isolation Scheme F: the board
,” that asked you to perform the Analog
output hardware test.
Table C-3
Test connections and correct readings for zero-voltage analog output, using an STA-300
screw terminal accessory connected to J1
To test this
analog output…
…connect the DVM or DMM to
these terminals on an STA-300
accessory:
If board works correctly, the
following voltages should agree:
Analog output
screw terminal
Analog-ground
screw terminal
Level control
setting
Voltage reading
at DVM or DMM
Analog output 0
[pin 58]
DAC0 Output
TB19
TB18
[pin 25]
0.0V
0.0V
Analog output 1
[pin 24]
DAC1 Output
TB23
0.0V
0.0V
Table C-4
Test connections and correct readings for mid-range analog output, using an STA-300
screw terminal accessory connected to the upper “Analog” I/O connector
To test this
analog output…
…connect the DVM or DMM to
these terminals on an STA-300
accessory:
If board works correctly, the
following voltages should agree:
Analog output
screw terminal
Analog-ground
screw terminal
Level control
setting
Voltage reading
at DVM or DMM
Analog output 0
[pin 58]
DAC0 Output
TB19
TB18
[pin 25]
5.0V
5.0V
Analog output 1
[pin 24]
DAC1 Output
TB23
5.0V
5.0V
Содержание KPCI-3101 Series
Страница 1: ...User s Manual KPCI 3101 3102 3103 3104 Series PCI Bus Data Acquisition Boards 98150 Rev A 10 99...
Страница 10: ...iv...
Страница 15: ...1 Overview...
Страница 21: ...2 Principles of Operation...
Страница 53: ...3 Installation and Configuration...
Страница 78: ...3 26 Installation and Configuration KPCI 3101 KPCI 3104 Series User s Manual...
Страница 79: ...4 Testing the Board...
Страница 82: ...4 4 Testing the Board KPCI 3101 KPCI 3104 Series User s Manual...
Страница 83: ...5 Calibration...
Страница 86: ...5 4 Calibration KPCI 3101 KPCI 3104 Series User s Manual...
Страница 87: ...6 Troubleshooting...
Страница 94: ...6 8 Troubleshooting KPCI 3101 KPCI 3104 Series User s Manual...
Страница 95: ...A Specifications...
Страница 107: ...B Connector Pin Assignments...
Страница 111: ...C Systematic Problem Isolation...
Страница 145: ...This page intentionally left blank...
Страница 146: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...