LAN reset
LAN RESET resets the LAN settings and the instrument password to the factory default values.To do this reset,
insert a straightened paper clip into the hole labeled LAN RESET.
LAN RESET is on the rear panel on the instruments. See figure below.
Figure 1: LAN Reset location
Sanitize instrument data
The only way to sanitize data from an instrument is to physically remove the nonvolatile memory devices listed
in the tables above.
Sanitize a nonfunctional instrument
To sanitize a nonfunctional instrument, remove the Digital and Display boards and return the instrument to
Keithley Instruments for installation of replacement boards.