Clearing data 

You can use TSP scripts to clear data. You can run these scripts from the front panel using a flash drive or run 

them remotely. 

Clearing data scripts and setups 

To clear the data scripts and setups: 

1.  Set the correct command set: 

*LANG TSP 

2.  Erase user strings from nonvolatile memory: 

for name in userstring.catalog() do 
   userstring.delete(name) 
end 

3.  Erase scripts and configuration lists from nonvolatile memory: 

for name in script.user.catalog() do 
   script.delete(name) 
end 

4. 

Source-measure instruments only (2450, 2460, 2461, and 2470):

 Set the GPIB address to the factory 

default: 

gpib.address = 18 

5.  Digital multimeters only (DMM7510): Set the GPIB address to the factory default: 

gpib.address = 16 

6.  Set the command set to SCPI: 

*LANG SCPI 

 

Resetting the instrument 

These steps only reset the volatile memory. Resetting the volatile memory restores the instrument to the factory 

defaults. 

If you are using the SCPI command set: 

*RST 

If you are using the TSP command set: 

reset() 

 