Model 2450 Interactive SourceMeter® Instrument User's Manual
Section 6: Leakage current and insulation resistance
2450-900-01 Rev. C / December 2013
6-5
7. Select the button next to Source (at the bottom of the screen).
8. Enter
20 V
and select
OK
.
9. Press the
MENU
key.
10. Under Trigger, select
Templates
.
11. Next to Templates, select
Duration Loop
.
12. Next to Duration, enter a soak time of
30s
and select
OK
.
13. Next to Delay, enter
0.2s
, and select
OK
.
14. Press the
HOME
key to return to the Home screen.
15. Press the
TRIGGER
key to turn on the output and make measurements. The output is turned off
when the measurements are complete.
View the measurements on the front-panel graph
To view the leakage current measurements on the front-panel graph:
1. Press the
MENU
key.
2. Under Views, select
Graph
.
3. Select the
Scale
tab.
4. Under Y-Axis, next to Scale Format, select
Log
.
5. Under X-Axis, next to Auto Scale, select
On
.
6. Select the
Graph
tab to view the graph.
The following figure shows the front-panel graph for this application.
Figure 27: Viewing the leakage current measurement on the front-panel