Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 2: Performance verification
DMM7510-905-01 Rev. A / April 2015
2-61
Verify the capacitance
Description
Verification point
Lower limit (F)
Upper limit (F)
1 nF range cable REL
3.00E-10
10 % 1 nF range
1.0000E-10
9.7000E-11
1.0300E-10
70 % 1 nF range
7.0000E-10
6.9100E-10
7.0900E-10
10 %10 nF range
1.0000E-09
9.8000E-10
1.0200E-09
100 % 10 nF range
1.0000E-08
9.8900E-09
1.0110E-08
10 % 100 nF range
1.0000E-08
9.8600E-09
1.0140E-08
100 % 100 nF range
1.0000E-07
9.9500E-08
1.0050E-07
10 % 1 µF range
1.0000E-07
9.8600E-08
1.0140E-07
100 % 1 µF range
1.0000E-06
9.9500E-07
1.0050E-06
10 % 10 µF range
1.0000E-06
9.8600E-07
1.0140E-06
100 % 10 µF range
1.0000E-05
9.9500E-06
1.0050E-05
10 %100 µF range
1.0000E-05
9.8600E-06
1.0140E-05
100 %100 µF range
1.0000E-04
9.9500E-05
1.0050E-04
10 % 1 mF range
1.0000E-04
9.8500E-05
1.0150E-04