Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 2: Performance verification
DMM7510-905-01 Rev. A / April 2015
2-41
2. Source
909.09
MΩ
.
3. On the reference DMM, enable
2-wire resistance
.
4. Enable the
2 GΩ
range.
5. Select
Res1
.
6. Set to
RESL '7'
.
7. Verify
909 MΩ
.
8.
On the reference DMM, record the 909 MΩ reading. Use this reading as the target for Model
DMM7510 1 GΩ verification.
Verifying 4-wire resistance accuracy
To verify 4-wire resistance accuracy:
1. Connect the Model DMM7510 INPUT and SENSE terminals to the calibrator as shown in the
following figure.
2. Set the calibrator for 4-wire resistance with external sense on.
3. On the Model DMM7510, press the
FUNCTION
key and select
4W Res
.
4. Press the
MENU
key.
5. Select Measure
Settings
.
6. Set Offset Compensation
On
.
7. Verify that Open Lead Detector is
Off
.