Section 3: Installation
Model 9139B Probe Card Adapter Instructions
3-10
9139B-901-01 Rev. A / January 2019
4. Align pin 1 of the probe card with pin 1 of the motherboard. Use the alignment marking on
the edge of the probe card retainer cap and match to the same markings of the motherboard
support ring.
5. Attach the retainer cap to the motherboard ring. Make sure latches secure the probe ring (retainer
cap will click into place).
6. Use the system diagnostics tool to check Kelvin connections from the system matrix to the probe
card adapter.
Probe card removal procedure
Hazardous voltages may be present on an installed probe card even after the output is
disconnected, that if contacted, may cause personal injury or death.
Never attempt to touch or change a probe card when tests are running. You must be
absolutely certain that all tests have stopped before making contact with anything in the
vicinity of the probe card adapter. Also, never run tests without a probe card installed.
To remove the probe card:
1. Turn the vacuum control box valve to the unload position.
2. Release the probe card retainer cap latches.
3. Carefully remove the probe card and retainer cap from the probe card adapter.