Rev. A
GENERAL
40 CHANNELS:
40 channels of 2-pole relay input.
All channels configurable to 4-pole.
2 CHANNELS:
2 channels of current only input.
RELAY TYPE:
Latching electromechanical.
ACTUATION TIME:
<3ms.
CAPABILITIES
CHANNELS 1-40:
Multiplex one of 40 2-pole or one of 20 4-pole signals
into DMM.
CHANNELS 41-42:
Multiplex one of 2 2-pole current signals into DMM.
INPUTS
MAXIMUM SIGNAL LEVEL:
Channels (1-40):
300V DC or rms, 1A switched, 60W, 125VA maximum.
Channels (41-42):
60V DC or 30V rms, 3A switched, 60W, 125VA maxi-
mum.
CONTACT LIFE (typ):
>10
5
operations at max signal level.
>10
8
operations cold switching.
CONTACT RESISTANCE:
<1
Ω
at end of contact life.
CONTACT POTENTIAL:
<±500nV typical per contact, 1µV max.
<±500nV typical per contact pair, 1µV max.
OFFSET CURRENT:
<100pA.
CONNECTOR TYPE:
Screw terminal, #20 AWG wire size.
ISOLATION BETWEEN ANY TWO TERMINALS:
>10
10
Ω
, <100pF.
ISOLATION BETWEEN ANY TERMINAL AND EARTH:
>10
9
Ω
, <200pF.
CROSS TALK (10MHz, 50
Ω
Load):
<–40dB.
INSERTION LOSS (50
Ω
Source, 50
Ω
Load):
<0.1dB below 1MHz.
<3dB below 2MHz.
COMMON MODE VOLTAGE:
300V between any terminal and chassis.
ENVIRONMENTAL
OPERATING ENVIRONMENT:
Specified for 0°C to 50°C.
Specified to 80% R.H. at 35°C.
STORAGE ENVIRONMENT:
–25°C to 65°C.
WEIGHT:
0.5kg (1.1 lb).
Specifications subject to change without notice.
Channel 1
HI
LO
Channel 20
HI
LO
(Channels 2–19)
Channel 21
HI
LO
Channel 40
HI
LO
(Channels 22–39)
Channel 41
HI
LO
Channel 42
HI
LO
AMPS
HI
LO
Sense
HI
LO
Input
Channel 43
2-Pole (Open)
4-Pole (Closed)
(see Note)
3A
3A
To Model 2700
Backplane
Channel 45
(see Note)
Backplane
isolation
Channel 44
(see Note)
Backplane
isolation
NOTE
Channels 43–45 in this schematic
refer to the designations used for
control and not actual available channels.
For more information, refer to the
ROUTe:MULTiple command section
in the Model 2700 User’s Manual.
Channels 44 and 45 can be individually
controlled using ROUTe:MULTiple if the
module is not to be connected to the
internal DMM.
Card Sense
HI
LO
Card Input
HI
LO
7702 40-Channel Differential Multiplexer
Содержание 7702