18
Model 7701
Figure 13
Jumper installation for common-side 4-wire ohms test system
Test procedure:
NOTE
The following test procedure assumes a Model 7701 switching module installed in slot 1 of the
mainframe.
Multiple channel operation from the front panel is not available for early versions of the Model
2700.
The following steps to test DUT in the common-side 4-wire ohm test system assumes that the
Ω
4 function of
the Model 2700/2750 is selected:
1.
Open all channels.
Front panel operation: Press OPEN > Display ALL > Press OPEN
Remote programming:
ROUT:OPEN:ALL
2.
Close channel 35 to connect measurement channels to DMM.
Front panel operation: Press CLOSE > Select MULTI > Key in 135 >
Press ENTER
Remote programming:
ROUT:MULT:CLOS (@135)
3.
Close channel 1 to test DUT1.
Front panel operation: Press CLOSE > Select MULTI > Key in 101 >
Press ENTER
Remote programming:
ROUT:MULT:CLOS (@101)
4.
Measure DUT #1.
Front panel operation: Take reading from display
Remote programming:
DATA? (for continuous triggering mode)
READ? (for one-shot triggering mode)
5.
Open channel 1 to disconnect DUT1 from DMM.
Front panel operation: Press OPEN > Select MULTI > Key in 101 >
Press ENTER
Remote programming:
ROUT:MULT:OPEN (@101)
6.
In general, repeat steps 3, 4, and 5 to test DUT2 through DUT32. That is, close the DUT measure-
ment channel, take a measurement, and then open the channel.
7.
After all DUT is tested, repeat step 1 to open all channels.
DMM HI
MUX 1 HI
DMM SHI
MUX 2 HI
DMM SLO
MUX 2 LO
DMM LO
MUX 1 LO
A
B
C
D
Jumper A
Jumper
(1 of 4)
P12
P13
P24
P25
Note: These screw teminal jumper connections are
required for the common-side 4-wire ohms test
system shown in Figure 12.
Jumper B
Jumper C
Jumper D
Connects the DUT to DMM Sense LO
through the switching channels.
Connects the DUT common-side
bus to DMM Input HI.
Connects the DUT common-side
bus to DMM Sense HI.
Connects DUT to DMM Input LO
through the switching channels.