Service Information
4-5
4.5
Special handling of static-sensitive
devices
CMOS and other high-impedance devices are subject to pos-
sible static discharge damage because of the high-impedance
levels involved. When handling such devices, use the precau-
tions listed below.
NOTE
In order to prevent damage, assume that all
parts are static sensitive.
1. Such devices should be transported and handled only in
containers specially designed to prevent or dissipate
static build-up. Typically, these devices will be received
in anti-static containers made of plastic or foam. Keep
these parts in their original containers until ready for in-
stallation or use.
2. Remove the devices from their protective containers
only at a properly-grounded work station. Also ground
yourself with an appropriate wrist strap while working
with these devices.
3. Handle the devices only by the body; do not touch the
pins or terminals.
4. Any printed circuit board into which the device is to be
inserted must first be grounded to the bench or table.
5. Use only anti-static type de-soldering tools and
grounded-tip soldering irons.
4.6
Performance verification
The following paragraphs discuss performance verification
procedures for the Model 7174A, including relay testing,
contact resistance, contact potential, path isolation, and leak-
age current.
4.6.1 Environment conditions
All verification measurements except for path isolation and
offset current should be made at an ambient temperature of
21°C-25°C and a relative humidity of less than 0% R.H. Path
isolation and offset current verification must be performed at
an ambient temperature of 23°C and a relative humidity of
less than 60% R.H. If the card has been subjected to temper-
atures or humidities outside of this range for even a short
time, allow it to stabilize within these ranges for at least one
hour before performing any tests.
4.6.2 Recommended test equipment
Table 4-3 summarizes the equipment necessary to make the
performance verification tests, along with the application for
each item.
Table 4-3
Recommended verification equipment
Qty.
Description
Application
1
1
1
4
2
1
3
5
1
1
1
Model 617 Electrometer
Model 196 6
½
Digit DMM
Model 707A Switching Matrix
Model 7078-TRX-10 triax cables*
Model 7078-TRX-3 triax cables
Model 6172 2-slot male to 3-lug female triaxial adapter
Model 7078-TRX-T triax tee adapter
Banana plugs (part #BG-10-2*)
Model 263 Calibrator/Source
BNC to Right-angle SMB Cable (part #CA-93-1)
BNC to Dual Banana Adapter (Pomona part #1269)
Offset current; path isolation
Path resistance; electrometer
All tests
Offset current; path resistance
Path isolation; offset current; electrometer
Offset current; electrometer
Path resistance
Path isolation and resistance
Electrometer
Electrometer
Electrometer
*These items are used to construct special cables; see text and Figure 4-9.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 7174A
Страница 68: ...M 1 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 69: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 70: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 71: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 72: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 73: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 74: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...