Models 707B and 708B Switching Matrix Reference Manual
Section 2: General operation
707B-901-01 Rev. A / August 2010
2-83
Change the instrument date and time
To change the date and time of the instrument:
1. In the web interface, from the left navigation, click
Admin
. A login page is displayed.
2. Enter the current password in the Password box (the default is "admin").
3. Click
Login
. The Administration page is displayed.
4. Enter the Year.
5. Select the Month, Day, Hour, Minutes and Seconds from the lists.
6. Click
Submit
. The new time and date information takes effect immediately.
Unit page
Create Config Script
: Save the set up of the instrument as a script.
To create the script:
1. Click
Create Config Script
. The Create Config Script dialog box is displayed.
2. To create a script that will run automatically when the instrument is powered on, select Auto-
execute on powerup. Note that this will overwrite the existing autoexec script.
3. To create a script with a new name, select Name and enter the name.
4. Click
OK
.
Reset
: Resets all instruments in the TSP-enabled system. This is only available if the instrument is
the master.
Open All
: Click this to open all relays on all slots.
Upgrade Firmware
: Select a firmware upgrade file to download to the instrument and begin the
upgrade process.
Channel Connect Rule
: Select the channel connect rule. See
(on page 2-89) for detail.
Digital I/O Lines
: This is the tool to configure the 14 digital I/O lines of the instrument. Values can be
read or written to the ports, or each individual bit can be toggled. "Write Protect" can be set
individually for any I/O line.