Section 3: Using the web interface
Models 707B and 708B 199BSwitching Matrix User's Manual
3-8
707B-900-01 Rev. B / June 2013
Exercise: Close channels in "Test1Pattern" from the web interface
To close channels in channel pattern "Test1Pattern" from the web interface:
1. Click
Pattern
(above the Channel Action Type box). The Channel Pattern Configuration dialog
box is displayed.
2. Select
Test1Pattern
from the list at the top of the dialog box.
In the Actions area, select the type of close you want to perform.
Figure 21: Close channels in a pattern
3. Click
Close
to close the channels.
4. When you are done acting on this channel pattern, click
Close
at the bottom of the dialog box to
exit.
To delete a channel pattern from the web interface:
1. Select the name of the pattern that you want to delete in the Channel Pattern Configuration dialog
box.
2. Click
Delete
.
3. For more information about patterns, including opening and closing the channels that are in
patterns, see
(on page 2-10).
Scan Builder page
The Scan Builder page allows you to set up and run scans and triggers.
A scan is a series of steps that opens and closes switches sequentially for a selected group of
channels. During each step, actions occur, such as waiting for a trigger, taking a measurement on an
external instrument, and completing a step count. Scans automate actions that you want to perform
consistently and repeatedly on a set of channels.
Triggers are events that prompt the instrument to move from one step to another in a scan. Triggers
can come from a variety of sources, such as a key press, digital input, or expiration of a timer. The
sequence of actions and events that occur during the scan is called the trigger model, described in
Trigger model.
Scanning and triggering allow you to synchronize actions across channels. You can set up a scan
using the trigger model to precisely time and synchronize the Model 707B or 708B between channels
and multiple instruments. You can also use triggers without the triggering model to set up a scan to
meet the needs of a specific application that does not fit the triggering model.
If you use Scan Builder to create a scan, use the options in the Scan Builder page to run the scan.
Using the TSB Embedded page may not give you the expected results.