Multiplexer Configuration
2-4
Differential switching
The differential or floating switching configuration is shown
in Figure 2-7. The advantage of using this configuration is
that the terminals of the source or measure instrument are not
confined to the same pathway. Each terminal of the instru-
ment can be switched to any available input in the test system.
Sensing
Figure 2-8 shows how the multiplexer can be configured to
use instruments that have sensing capability. The main
advantage of using sensing is to cancel the effects of switch
card path resistance (<1.25
Ω
) and the resistance of external
cabling. Whenever path resistance is a consideration, sensing
should be used.
Figure 2-7
Differential switching example
Figure 2-8
Sensing example
HI
LO
Source or
Measure
Bank A
7021
H
L
H
L
DUT
Out
Out
In
In
Bank B
Source HI
Sense HI
Sense LO
Source LO
Source or
Measure
Bank A
7021
H
L
H
L
DUT
Out
Out
Bank B
In
In
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Содержание 7021
Страница 63: ...Operation 5 16 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 86: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 87: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 88: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 89: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 90: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...