2-4
2.3.5 Input termination
User-supplied 50
Ω
female SMB terminators may be in-
stalled for applications requiring proper 50
Ω
termination of
off channels. Figure 2-4 shows typical installation of a 50
Ω
terminator. See paragraph 1.10 in Section 1 for recommend-
ed terminators. Switching considerations in Section 3 dis-
cusses terminator aspects in more detail.
NOTE
Be sure to observe the maximum power
handling capability of installed termina-
tors. The terminators recommended in
Section 1 are limited to 2W even through
the Model 7016A can switch up to 10W.
Terminator
Jack
Figure 2-4
50
Ω
terminator installation
2.4
Card installation and removal
This paragraph explains how to install and remove the Model
7016A card assembly from the Model 7001/7002 main-
frame.
WARNING
Installation or removal of the Model
7016A should be performed only by
qualified service personnel. Failure to
recognize and observe standard safety
precautions could result in personal in-
jury or death.
NOTE
To prevent performance degradation
caused by contamination, handle the card
only by the edges and covers.