4.3.4 Offset current tests
Offset current tests are performed by measuring the residual
current of each Model 6521/22 channel by closing each
channel in succession and then measuring the offset current
using the Model 6517 Electrometer.
Proceed as follows:
1. With the power off, install the Model 6521/22 Scanner
Card in the Model 6517 OPTION SLOT.
2. Connect the scanner card triax output cable to the Model
6517 INPUT jack. (See Figure 4-1.) Also install the
BNC (Model 6521) or triax (Model 6522) shielding cap
on each channel 1 input cable connector.
3. Program the Model 6517 for the current scan mode (see
paragraph 3.4.4).
4. Turn on the electrometer, and allow a one-hour warm-up
period before performing the offset current tests.
5. Select the electrometer amps function and the 20pA
range. Leave zero check enabled for the time being.
6. Press the OPTION CARD button, then use the
CHANNEL-CLOSURES menu to close channel 1 on
the scanner card.
7. Disable zero check, and allow the reading to settle.
8. Verify that the Model 6517 current reading, which rep-
resents the scanner card offset current, is <1pA.
9. Enable zero check, then open channel 1.
10. Repeat steps 6 through 9 for channels 2 through 10. For
each channel, be sure to connect the shielding cap to the
input cable, close the appropriate channel, then verify
that the offset current is <1pA.
Figure 4-1
Connections for offset current verificatio
Model
6517
Electrometer
Triax Cable
Install
Shielding
Caps on
Inputs
H
L IN 1
H
L IN 10
IN 2....9
H
L
G
OUT
HI
LO
Ground
Model 6521
Note: Model 6522 connection
is the same except for
triax input.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
INPUT
250V PEAK
!
LINE RATING
50-60HZ
50VA MAX
AC ONLY
LINE FUSE
SLOWBLOW
1/2A 90-125V
1/4A 180-250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
DIGITAL
I/O
TRIG LINK
115V
Содержание 6521
Страница 12: ...iv...
Страница 14: ...vi...
Страница 16: ...viii...
Страница 20: ...General Information 1 4...
Страница 28: ...Card Connections and Installation 2 8...
Страница 42: ...Operation 3 14...
Страница 56: ...Replaceable Parts 5 4...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 63: ......
Страница 64: ......