Card Connections and Installation
2-3
2.3.3 Input connections
All connections to the scanner card input jacks should be
made using appropriate low-noise coax (Model 6521) or tri-
ax (Model 6522) cables. See paragraph 1.9 in Section 1 for
cable recommendations.
WARNING
Make sure all power is off and any
stored energy in external circuitry is dis-
charged before connecting or discon-
necting cables.
2.3.4 Output connections
The supplied triax output cable should be connected to the
Model 6517 INPUT jack.
2.3.5 Typical connecting schemes
Figure 2-5 through Figure 2-8 show typical connecting
schemes for using the scanner card for amps, volts, ohms,
and coulombs measurements.
NOTE
The Model 6521 is intended for current
measurements. The Model 6522 is de-
signed for all Model 6517 measurement
functions. The Model 6517 should be pro-
grammed for the voltage or current scan
mode as appropriate. (See paragraph
3.4.4.)
Figure 2-4
Model 6522 connecting cables
Triax Input
cables
Triax output
cable
Figure 2-5
Typical amps function connections
To
6517
Input
Jack
Current
Sources (10)
H
Out L
G
H
L
IN 2....9
H
L
1
10
Note: Model 6522 similar
exept that triax inputs
are used.
Coax
Input
Cables
Model 6521
Figure 2-6
Typical volts function connections
To
6517
Input
Jack
Voltage
Sources (10)
H
Out L
G
H
L
IN 2....9
H
L
1
10
Triax
Input
Cables
G
G
Model 6522
V
V
Содержание 6521
Страница 12: ...iv...
Страница 14: ...vi...
Страница 16: ...viii...
Страница 20: ...General Information 1 4...
Страница 28: ...Card Connections and Installation 2 8...
Страница 42: ...Operation 3 14...
Страница 56: ...Replaceable Parts 5 4...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 63: ......
Страница 64: ......