2-28
Measurement Options
Model 6517A Getting Started Manual
Close/Open channels
Perform the following steps to open or close a channel on an internal scanner card:
1.
Press the CARD key to display the internal scanner options (close channel or per-
form scan).
2.
Place the cursor (using
and
) on CHANNEL-CLOSURES and press ENTER.
3.
The following channel closure options are available:
a.
Close channel — To close a channel, place the cursor on CLOSE-CHANNEL
and press ENTER. Use the RANGE keys to display the channel that you wish
to close and press ENTER. The number of the closed channel will be dis-
played along with the reading.
b.
Open channels — To open all channels on the scanner card, place the cursor
on OPEN-ALL-CHANNELS and press ENTER.
External scanning
The Model 6517A can be used with a scanner card (such as the Model 7158 Low Current
Multiplexer Card) installed in an external scanning mainframe (Model 706, 7001, or 7002
Switch System). With the use of external triggering, the Model 6517A can measure and
store each scanned channel.
NOTES
For complete information on scanning (external and internal), refer to Section 2
in the User’s Manual, and the switch system and scanner card instruction
manual.
The following steps assume that the Model 6517A is set to the BENCH reset
default conditions (see Section 2, “Bench Default”) and the Model 7001/7002 is
set to the RESET default conditions.
Trigger connections
If using Trigger Link, connect the Model 6517A to the switch system as shown in
Figure 2-8. If using conventional external triggering, connect the instruments as shown in
Figure 2-9. Detailed information on triggers is provided in Section 2 of the User’s Manual.