SCPI Command Reference
17-59
Set compliance limit
[:LEVel] <n>
[:SENSe[1]]:CURRent[:DC]:PROTection[:LEVel] <n>
Set current compliance
[:SENSe[1]]:VOLTage[:DC]:PROTection[:LEVel] <n>
Set voltage compliance
Parameters
<n> =
-105e-3 to 105e-3
Current compliance limit
-210 to 210
Voltage compliance limit
DEFault
105uA, 21V
MINimum
-105e-3A, -210V
MAXimum
105e-3A, 210V
Query
:LEVel?
Query compliance value
:LEVel? DEFault
Query *RST default compliance
:LEVel? MINimum
Query minimum allowable compliance
:LEVel? MAXimum
Query maximum allowable compliance
Description
This command is used to set compliance limits. A current compliance limit
is set for the V-Source, and a voltage compliance limit is set for the I-
Source. The SourceMeter cannot source levels that exceed these speci
fi
ed
limits.
The :SENSe:CURRent:PROTection[:LIMit] command is used to set the
current compliance for the V-Source and the
:SENSe:VOLTage:PROTection[:LIMit] command is used to set the voltage
compliance for the I-Source.
NOTE
You cannot set compliance less than 0.1% of the present measurement range.
:TRIPped?
[:SENSe[1]]:CURRent[:DC]:PROTection:TRIPped?
Query current compliance state
[:SENSe[1]]:VOLTage[:DC]:PROTection:TRIPped?
Query voltage compliance state
Description
This command is used to determine if the source is in compliance. If a
“
1
”
is
returned, then the source is in compliance. A
“
0
”
indicates that the source is
not in compliance.
The :CURRent:PROTection:TRIPped? command is used to check the com-
pliance state of the V-Source, and the :VOLTage: PROTection:TRIPped?
command is used to check the compliance state of the I-Source.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......