4-10
Sweeps
Model 6220/6221 Reference Manual
Return to
Front panel sweep operation
Using the sweep configuration menu
To configure sweeps, press CONFIG then SWP, and then make your selections
from
below. See the detailed procedures for each sweep type in the fol-
lowing paragraphs.
NOTE
User setups cannot be saved or recalled while a
sweep is armed or running. Attempting to do so
will generate error +413 Not allowed with mode
armed.
Table 4-2
Sweep configuration menu
Menu selection
Description
TYPE
STAIR
LOG
CUSTOM
Select sweep type:
Set START, STOP, STEP, DELAY.
Set START, STOP, NO OF POINTS, DELAY.
Set #-POINTS, ADJUST POINTS, AUTO COPY.*
SWEEP-COUNT
FINITE
INFINITE
Choose sweep count:
Enter desired # of sweeps.
Continuously repeating sweeps.
SOURCE-RANGING
BEST
AUTO
FIXED
Select source ranging:
Use best range based on maximum sweep step.
Auto range based on individual sweep step.
Always stay on fixed source range.
COMPLIANCE-ABORT
NO
YES
Select compliance abort mode:
Do not abort sweep if compliance reached.
Abort sweep if compliance reached.
* Select AUTO COPY ON to automatically copy delay and compliance values to all sweep points.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...