![Keithley 4210-MMPC-C Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/keithley/4210-mmpc-c/4210-mmpc-c_quick-start-manual_3938553002.webp)
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
2
PA-1001 Rev. D / September 2011
Model 4210-MMPC-C
Multi-Measurement Prober Cable Kit
Quick Start Guide
Quick Start Guide topics
The guide contains information about the installation and use of this cable kit:
Connecting Model 4200
instrumentation to the prober bulkhead: Explains how to make connections from
the Model 4200 to the prober bulkhead.
Installing the prober cable kits
: Explains how to make connections from the prober bulkhead to the prober
pins.
Usage scenarios:
Provides the specific prober cable setups:
o
I-V testing (2-wire and 4-wire)
o
C-V testing (2-wire and 4-wire)
o
Pulse I-V measurements using the Keithley Model 4200-PIV-A
o
Pulse I-V measurements using the Keithley Model 4225-PMU with the 4225-RPM
Related documents
The following documents (in PDF format) are located on the Model 4200 Complete Reference CD:
PA-993:
This is a one-page packing list for the Model 4210-MMPC-S Multi-Measurement Prober Cable Kit
for the
Suss MicroTec PA200/300 series
prober. A photo shows all the components of the cable kit.
PA-994:
This is a one-page packing list for the Model 4210-MMPC-C Multi-Measurement Prober Cable Kit
for the Cascade MicroTech 12000 series prober. A photo shows all the components of the cable kit.
Section 16
of the 4200-SCS Reference Manual – The Reference Manual PDF can be found on the main
4200-SCS Complete Reference. Open the Complete Reference web page by double-clicking the Complete
Reference icon in the upper-left corner of the 4200 desktop.
White Paper
– Labs’ Demands for Greater Measurement Flexibility Require Cabling Systems Capable of
Accommodating Multiple Measurement Types: This white paper explains the different cabling requirements
for I-V, C-V, and pulse I-V testing. It also describes an easy-to-use single multi-measurement cabling
system (Model 4210-MMPC-C for the Cascade Microtech 12000 prober).
Connecting Model 4200 instrumentation to the prober bulkhead
WARNING
To prevent injury or death due to electric shock, remove all power from the
Model 4200-SCS and shut down the Cascade prober before installing the
prober cable kit.
The Model 4200 connects to the female triax connectors on the prober bulkhead.
I-V testing
For I-V testing, connect the triax cables of your Model 4200 source-measure units (SMUs) to the standard female
triax connectors on the bulkhead of the prober. Figure 2 shows two Model 4200 SMU prober connections.