4-4
Range, Digits, Rate, Bandwidth, and Filter
Model 2750 Multimeter/Switch System User’s Manual
Remote programming — range
Range commands
The commands to set range are listed in
. Additional information on these com-
mands follow the table.
NOTE
Query commands and some optional command words are not included in
. All commands for the SENSe subsystem are provided in
Table 4-2
Range commands
Commands
1, 2
Description
Default
[SENSe[1]]
Optional root command.
:VOLTage[:DC]:RANGe <n> [, <clist>]
Select DCV range; <n> = 0 to 1010 (V).
1000
:VOLTage[:DC]:RANGe:AUTO <b>
[, <clist>]
Control DCV auto range; <b> = ON or OFF. ON
:VOLTage:AC:RANGe <n> [, <clist>]
Select ACV range; <n> = 0 to 757.5 (V).
750
:VOLTage:AC:RANGe:AUTO <b> [, <clist>]
Control ACV auto range; <b> = ON or OFF. ON
:CURRent[:DC]:RANGe <n> [, <clist>]
Select DCI range; <n> = 0 to 3 (A).
3
:CURRent[:DC]:RANGe:AUTO <b>
[, <clist>]
Control DCI auto range; <b> = ON or OFF.
ON
:CURRent:AC:RANGe <n> [, <clist>]
Select ACI range; <n> = 0 to 3 (A).
3
:CURRent:AC:RANGe:AUTO <b> [,
<clist>]
Control ACI auto range; <b> = ON or OFF.
ON
:RESistance:RANGe <n> [, <clist>]
Select
Ω
2 range; <n> = 0 to 120e6 (
Ω
). 120e6
:RESistance:RANGe:AUTO <b> [, <clist>]
Control
Ω
2 auto range; <b> = ON or OFF.
ON
:FRESistance:RANGe <n> [, <clist>]
Select
Ω
4 range; <n> = 0 to 120e6 (
Ω
).
120e6
:FRESistance:RANGe:AUTO <b> [, <clist>]
Control
Ω
4 auto range; <b> = ON or OFF.
ON
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
1. The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the range command. If not, a conflict error (-221) will occur. For example, VOLTage:AC:RANGe 1, (@101) is only
valid if scan channel 101 is set for the ACV function.
2. [:DC] is optional for the commands to set DCV and DCI range.
2750-900-01.book Page 4 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...