Model 2657A-PM-200 Protection Module 

 

  

PA-1055 Rev. A / February 2012 

 

Connections using a Model 4200-SCS SMU (with Model 4200-PA) 

If ordered, a Model 4200-PA remote preamplifier (preamp) for a source-measure unit (SMU) is installed at the 
factory when the Model 4200-SCS is ordered.   

Required accessories to connect one Model 4200-SMU or 4210-SMU (with the Model 4200-PA): 

 

Two or three Model 4200-TRX cables 

 

Two or three high-voltage (HV) triaxial cables (Model CA-553 or CA-554) 

Connection summary 

Use the connection drawing below to connect a Model 4200-PA preamp and the GNDU (ground unit) to the Model 
2657A-PM-200. Make sure to use the IN connectors of the Model 2657A-PM-200 for connection to the Model 4200-
SCS and use the OUT connectors for connection to the device under test (DUT).  

The drawing shows connections for 4-wire sensing. For 2-wire applications, the SH (sense HI) connectors of the 
Model 2657A-PM-200 are not used.  

When using multiple SMUs that are installed in the Model 4200-SCS, each SMU needs its own 
protection module. To achieve adequate voltage protection, connect the GNDU Sense terminal to the 
LO,SL (LO and sense LO) connector of each protection module using triax tees (such as the Keithley 
Model 237-TRX-T) and additional triaxial cables (Model 4200-TRX). 

 

Connecting the Model 4200-SMU or 4210-SMU to the Model 2657A-PM-200 connects the LO 
terminal of the SMU to protective earth (safety ground).   

 

Do not convert triaxial cables to BNC cables. Using BNC cables removes protection from the Model 
4200-SCS SMU and may result in damage to the SMU. You must use triaxial cables to connect to the 
lower-voltage SMU.  

Содержание 2657A-PM-200

Страница 1: ...t Model 2635A Single Channel System SourceMeter Instrument Model 2636A Dual Channel System SourceMeter Instrument Model 4200 SCS Semiconductor Characterization System SMUs Model 4200 SMU Medium Power...

Страница 2: ...e 2 Model 2657A PM 200 grounding The standard triaxial IN connectors are used for connection to the SMU Maximum clamped voltage on the IN connectors is 240 VDC The high voltage triaxial connectors are...

Страница 3: ...m in unclamped condition Pulsed current carrying capability 10 A maximum at 1 duty cycle in unclamped condition Maximum protection active current User must limit current to 120 mA in clamped condition...

Страница 4: ...ions for 4 wire sensing For 2 wire applications the SH sense HI connectors of the Model 2657A PM 200 are not used Connecting the Model 2600 TRIAX to the Model 2657A PM 200 connects the LO terminal of...

Страница 5: ...Model 2657A PM 200 Protection Module PA 1055 Rev A February 2012 5 Figure 4 Connections for a Model 2611A or 2612A...

Страница 6: ...connection to the device under test DUT The drawing shows connections for 4 wire sensing For 2 wire applications the SH sense HI connectors of the Model 2657A PM 200 are not used Do not convert triax...

Страница 7: ...Model 2657A PM 200 Protection Module PA 1055 Rev A February 2012 7 Figure 5 Connections for a Model 2635A or 2636A...

Страница 8: ...d use the OUT connectors for connection to the device under test DUT The drawing shows connections for 4 wire sensing For 2 wire applications the SH sense HI connectors of the Model 2657A PM 200 are n...

Страница 9: ...al difference between the two circuits in excess of 100 mV may result in compromised measurements Large potential differences can be created when there is significant lead resistance and current flowi...

Страница 10: ...the Model 2657A PM 200 are not used When using multiple SMUs that are installed in the Model 4200 SCS each SMU needs its own protection module To achieve adequate voltage protection connect the GNDU S...

Страница 11: ...Model 2657A PM 200 Protection Module PA 1055 Rev A February 2012 11 Figure 7 Connections for a Model 4200 SCS SMU...

Страница 12: ...device is connected between Guard and LO Low leakage diodes are used to connect both HI and Sense HI to the Guard terminal This enables the protection module to protect a SMU connected to the IN term...

Страница 13: ...of the protection module a Connect the SMU Force HI terminal to the center conductor of the SH IN triaxial connector of the Model 2657A PM 200 b Connect the SMU Force LO terminal to the inner shield o...

Страница 14: ...nductor of the HI OUT triaxial connector of the Model 2657A PM 200 b Connect the SMU Force LO terminal to the inner shield of the HI OUT triaxial connector of the Model 2657A PM 200 Figure 10 Connecti...

Страница 15: ...must be rated to double the maximum capability of the test equipment in the system and the two green yellow ground cables for the Model 2657A PM 200 must be connected to protective earth safety groun...

Страница 16: ...ial connector of the protection module a Connect the SMU force HI terminal to the inner shield of the LO SL IN triaxial connector of the Model 2657A PM 200 b Connect the SMU force LO terminal to the o...

Страница 17: ...ng a nonconductive test fixture it must be rated to double the maximum capability of the test equipment in the system and the two green yellow ground cables for the Model 2657A PM 200 must be connecte...

Страница 18: ...off the HV SMU output 10 Connect the scope across the output terminals of the HV SMU 11 Configure the HV SMU to source voltage 12 Program the voltage source level to 300 V 13 Turn on the HV SMU outpu...

Страница 19: ...the Model 2657A PM 200 b Connect the SMU Force LO terminal to the inner shield of the LO SL OUT triaxial connector of the Model 2657A PM 200 Figure 14 Connections for verification procedure F 2 Config...

Страница 20: ...scope has a ramped waveform The voltage should ramp up to at least 150 V and then clamp to less than 50 V 15 Turn off the HV SMU output 16 Program the voltage source level to 300 V 17 Turn on the HV...