Model 2657A-PM-200 Protection Module 

 

  

PA-1055 Rev. A / February 2012 

 

Connections using a Model 2635A or 2636A SMU 

Required accessories: 

 

Two or three standard triaxial cables (either Model 7078-TRX or 4200-TRX) per source-measure unit (SMU) 
channel 

 

Two or three high-voltage (HV) triaxial cables (Model CA-553 or CA-554) per SMU channel 

Connection summary 

Use the connection drawing below to connect source-measure unit (SMU) channels of the Model 2635A or 2636A 
to the Model 2657A-PM-200. For each channel, make sure to use the IN connectors of the Model 2657A-PM-200 
for connection to the Model 2635A or 2636A and use the OUT connectors for connection to the device under test 
(DUT).  

The drawing shows connections for 4-wire sensing. For 2-wire applications, the SH (sense HI) connectors of the 
Model 2657A-PM-200 are not used.  

Do not convert triaxial cables to BNC cables. Using BNC cables removes protection from the Model 
2635A or 2636A and may result in damage to the SMU. You must use triaxial cables to connect to the 
lower-voltage SMU.  

When using the 2657A-PM-200, HI and SH (sense HI) should be connected together at the device 
under test. A potential difference between the two circuits in excess of 100 mV may result in 
compromised measurements. 

Large potential differences can be created when there is significant lead resistance and current 
flowing in the HI lead. To minimize lead resistance, minimize the distance between the Model 2657A-
PM-200 and the device under test.  

Содержание 2657A-PM-200

Страница 1: ...t Model 2635A Single Channel System SourceMeter Instrument Model 2636A Dual Channel System SourceMeter Instrument Model 4200 SCS Semiconductor Characterization System SMUs Model 4200 SMU Medium Power...

Страница 2: ...e 2 Model 2657A PM 200 grounding The standard triaxial IN connectors are used for connection to the SMU Maximum clamped voltage on the IN connectors is 240 VDC The high voltage triaxial connectors are...

Страница 3: ...m in unclamped condition Pulsed current carrying capability 10 A maximum at 1 duty cycle in unclamped condition Maximum protection active current User must limit current to 120 mA in clamped condition...

Страница 4: ...ions for 4 wire sensing For 2 wire applications the SH sense HI connectors of the Model 2657A PM 200 are not used Connecting the Model 2600 TRIAX to the Model 2657A PM 200 connects the LO terminal of...

Страница 5: ...Model 2657A PM 200 Protection Module PA 1055 Rev A February 2012 5 Figure 4 Connections for a Model 2611A or 2612A...

Страница 6: ...connection to the device under test DUT The drawing shows connections for 4 wire sensing For 2 wire applications the SH sense HI connectors of the Model 2657A PM 200 are not used Do not convert triax...

Страница 7: ...Model 2657A PM 200 Protection Module PA 1055 Rev A February 2012 7 Figure 5 Connections for a Model 2635A or 2636A...

Страница 8: ...d use the OUT connectors for connection to the device under test DUT The drawing shows connections for 4 wire sensing For 2 wire applications the SH sense HI connectors of the Model 2657A PM 200 are n...

Страница 9: ...al difference between the two circuits in excess of 100 mV may result in compromised measurements Large potential differences can be created when there is significant lead resistance and current flowi...

Страница 10: ...the Model 2657A PM 200 are not used When using multiple SMUs that are installed in the Model 4200 SCS each SMU needs its own protection module To achieve adequate voltage protection connect the GNDU S...

Страница 11: ...Model 2657A PM 200 Protection Module PA 1055 Rev A February 2012 11 Figure 7 Connections for a Model 4200 SCS SMU...

Страница 12: ...device is connected between Guard and LO Low leakage diodes are used to connect both HI and Sense HI to the Guard terminal This enables the protection module to protect a SMU connected to the IN term...

Страница 13: ...of the protection module a Connect the SMU Force HI terminal to the center conductor of the SH IN triaxial connector of the Model 2657A PM 200 b Connect the SMU Force LO terminal to the inner shield o...

Страница 14: ...nductor of the HI OUT triaxial connector of the Model 2657A PM 200 b Connect the SMU Force LO terminal to the inner shield of the HI OUT triaxial connector of the Model 2657A PM 200 Figure 10 Connecti...

Страница 15: ...must be rated to double the maximum capability of the test equipment in the system and the two green yellow ground cables for the Model 2657A PM 200 must be connected to protective earth safety groun...

Страница 16: ...ial connector of the protection module a Connect the SMU force HI terminal to the inner shield of the LO SL IN triaxial connector of the Model 2657A PM 200 b Connect the SMU force LO terminal to the o...

Страница 17: ...ng a nonconductive test fixture it must be rated to double the maximum capability of the test equipment in the system and the two green yellow ground cables for the Model 2657A PM 200 must be connecte...

Страница 18: ...off the HV SMU output 10 Connect the scope across the output terminals of the HV SMU 11 Configure the HV SMU to source voltage 12 Program the voltage source level to 300 V 13 Turn on the HV SMU outpu...

Страница 19: ...the Model 2657A PM 200 b Connect the SMU Force LO terminal to the inner shield of the LO SL OUT triaxial connector of the Model 2657A PM 200 Figure 14 Connections for verification procedure F 2 Config...

Страница 20: ...scope has a ramped waveform The voltage should ramp up to at least 150 V and then clamp to less than 50 V 15 Turn off the HV SMU output 16 Program the voltage source level to 300 V 17 Turn on the HV...