22-8
Return to
2600AS-901-01 Rev. B / September 2008
Section 22: Performance Verification
Series 2600A System SourceMeter® Instruments Reference Manual
•
Set the Series 2600A output voltage to the indicated value.
•
Verify that the multimeter reading is within the limits given in the table.
1.
Repeat the procedure for negative output voltages with the same magnitudes as those
listed in
.
2.
For the Model 2602A/2612A/2636A, repeat the above procedure for the other channel.
Voltage measurement accuracy
Follow the steps below to verify that the Series 2600A voltage measurement accuracy is within
specified limits. The test involves setting the source voltage, as measured by a precision digital
multimeter, and then verifying that the Series 2600A voltage readings are within required limits.
1.
With the power off, connect the digital multimeter to the Series 2600A output terminals using
4-wire connections, as shown in
.
2.
Select the multimeter DC volts function.
3.
Select the Model 2602A/2612A/2636A single-channel display mode.
4.
Enable the Series 2600A 4-wire (remote sense) mode by pressing
CONFIG
then
MEAS
,
then select
V-MEAS
>
SENSE-MODE
>
4-WIRE
.
5.
Set the Series 2600A to both source and measure voltage by pressing the
SRC
and
MEAS
keys, and make sure the source output is turned on.
6.
Verify voltage measurement accuracy for each of the voltages listed in
(Model
(Model 2611A/2612A/2635A/2636A). For each test point:
•
Select the correct source range.
•
Set the Series 2600A output voltage to the indicated value as measured by the digital
multimeter.
•
If necessary, press the
TRIG
key to display readings.
•
Verify that the Series 2600A voltage reading is within the limits given in the table. It may
not be possible to set the voltage source to the required value. Use the closest possible
setting, and modify reading limits accordingly.
7.
Repeat the procedure for negative source voltages with the same magnitudes as those
listed in
.
Table 22-2
Model 2601A/2602A output voltage accuracy limits
Model 2601A/2602A
source range
Model 2601A/2602A
output voltage setting
Output voltage limits
(1 year, 18°C–28°C)
100mV
1V
6V
40V
90.000mV
0.90000V
5.4000V
36.000V
89.732 to 90.268mV
0.89942 to 0.90058V
5.39712 to 5.40288V
35.9808 to 36.0192V
Table 22-3
Model 2611A/2612A/2635A/2636A output voltage accuracy limits
Model 2611A/
2612A/2635A/
2636A
source range
Model 2611A/2612A/
2635A/2636A output
voltage
setting
Output voltage limits
(1 year, 18°C–28°C)
200mV
2V
20V
200V
180.000mV
1.80000V
18.000V
180.000V
179.589 to 180.411mV
1.79904 to 1.80096V
17.9914 to 18.0086V
179.914 to 180.086V