2600AS-901-01 Rev. B / September 2008
Return to
22-7
Series 2600A System SourceMeter® Instruments Reference Manual
Section 22: Performance Verification
Figure 22-1
Connections for voltage verification
2.
Select the multimeter DC volts measuring function.
3.
Select the Model 2602A/2612A/2636A single-channel display mode. Press the
SRC
key to
source voltage and make sure the source output is turned on.
4.
Enable the Series 2600A 4-wire (remote sense) mode by pressing
CONFIG
then
SRC
, then
select
V-SOURCE
>
SENSE-MODE
>
4-WIRE
.
5.
Verify output voltage accuracy for each of the voltages listed in
(Model 2601A/
2602A) or
(Model 2611A/2612A/2635A/2636A). For each test point:
•
Select the correct source range.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
SENSE
LO
LO
HI
CHANNEL A
SENSE
HI
GUARD
A LO
LAN
NO AUTO-MDIX
SENSE
LO
LO
HI
CHANNEL B
SENSE
HI
GUARD
!
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
S
S
LO
LO
HI
HI
G G G
G
CHANNEL B
!
Input HI
Input LO
Model 2636A Channel A
Connections Shown
LO
HI
Digital Multimeter
S LO
S HI
Input HI
Input LO
Model 2602A/2612A Channel A
Connections Shown
LO
HI
Digital Multimeter
S LO