Model 2600-FIX-TRX 

Keithley Instruments, Inc.

 

 

28775 Aurora Road

 

Cleveland, Ohio 44139 
1-888-KEITHLEY 

www.keithley.com

 

 

Grounded Phoenix-to-Triax Cable Adapter 

PA-1004 Rev. A / June 2009 

 

Figure 2: Connector schematic 

 

 

Proper use of these cable adapters and fixtures together ensures that the source-measure units used to test a 
device share a common ground connection, yielding accurate measurements.  

NOTE 

The Model 2600-FIX-TRX should not be used for 4-wire measurements because the sense leads are not 
connected in these connector assemblies.  

 

Example test connections 

The example i

Figure 3

 demonstrates how to connect two Model 2602A System SourceMeter Instruments to a 

Model 8101-4TRX test fixture. 